10 Optional Functions
109
Q
Time 2
Pressing the button at the right of [TIME 2] in the ROLLING AVERAGE TIME SET screen
displays the ROLLING AVERAGE TIME SET 2 screen.
Fig. 84 ROLLING AVERAGE TIME SET 2 screen
Pressing the numerical keypad to input the value.
Pressing each key allows the corresponding operation as follows:
Note
Setting the ROLLING AVERAGE TIME 2 more than 240 minutes automatically sets the time to 240
minutes.
10.3.2 Integration time
The integration time for the integration value is fixed to each one hour.
It is reset per hour from power ON or from the integration value is reset.
Note
z
In purge, in calibration ,and in maintenance (with the output hold specification), the rolling
average time as well as the integration time are accumulated and calculated just before they are
held.
z
In changing the rolling average time, the data before the change is temporarily reset, and the time
newly changed is used for the calculation.
Setting item
Setting range
ROLLING AVERAGE TIME 1 to 240
CANCEL
Returns to the
ROLLING AVERAGE TIME SET
screen without changing the
setting.
CLEAR
The value input to the editing area is cleared.
BACK
The last digit input is cleared.
SET
Returns to the
ROLLING AVERAGE TIME SET
screen with the setting
changed.