10 Optional Functions
111
10.4 Measurement Value Alarm
When the measurement concentration is over or below the set value, the alarm can be
triggered.
10.4.1 Measure value alarm
This section describes the setting operation of measure value alarm.
There are three types for setting the alarm: HIGH, H.HIGH and LOW.
1.
Press the [MENU] key in the MEAS. screen.
Press the [
X
] key three times to display the MENU 4/5 (OPTION menu) screen.
Fig. 87 MENU 4/5 (OPTION menu) screen
2.
Press the [MEASURE VALUE ALARM] button.
The MEASURE VALUE ALARM screen appears.
Fig. 88 MEASURE VALUE ALARM screen