Maintenance
4-1
Caution:
The integrating sphere is deli-
cate. Mishandling may seriously dam-
age its interior reflective surface.
Caution:
Never use Freon
®
or any other organic solvent
to clean the Spectralon
®
surfaces. Always read the Ma-
terial Safety Data Sheets before using dry nitrogen or
other pressurized gases.
4
: Maintenance
Handling
Wear clean
gloves when handling the Integrating-Sphere halves or the optional calibrated reflector,
in order to prevent the skin’s natural oils from contaminating the Spectralon
®
surface.
Cleaning
Light soil
If the interior of the Integrating Sphere or the surface of the optional calibrated reflector
becomes lightly soiled, blow a jet of clean, dry air or nitrogen gas on the surface.
Heavy soil
Contact HORIBA Scientific for advice.