Quanta-
φ
rev. C (23 Apr 2010)
Installation & Use
3-25
Using a photomultiplier tube as the detec-
tor
Instead of a CCD detector, a photomultiplier tube (PMT) may be used as the detector.
Data-acquisition is slightly different in this case.
For a 2-curve analysis:
1
Place the blank into the integrating sphere, and
properly close the sphere.
2
Start FluorEssence™; run an emission scan.
3
In the
Experiment Setup
window, set the parame-
ters as shown below:
a
Enter a
Data Identifier
to indicate a blank scan.
b
In the above case, our PMT is on the T-side, so we de-activate the S-
detector (Symphony iHR) checkbox, and activate the
Emission 2 (T1)
checkbox. Your system may be different.