Quanta-
φ
rev. C (23 Apr 2010)
Installation & Use
3-12
h
Enable the emission detector, the
Correction
checkbox, and the
Dark
Offset
checkbox. Set the
Integration Time
to 0.1 s.
i
Add >>
the corrected signal (with
c
subscript) to the
Formulas
table.
j
Give this file a
Data Identifier
.
k
Choose
1
Accumulations
.
If this is too noisy, you will increase the accumulations later.
l
Click the
Run
button.
The
Interme-
diate Display
appears.
m
A
Warning
win-
dow may appear
that the
length range
cludes values
outside of the
correction-factor
file’s range.