Quanta-
φ
rev. C (23 Apr 2010)
Installation & Use
3-18
11
Set up the analysis of the data.
There are two methods of analyzing the data: 2-curve analysis and 4-curve
analysis. In the 2-curve analysis, the Rayleigh line and luminescence band are
included together in the sample’s and blank’s data. Thus only two curves are
used for analysis. In the 4-curve method, the Rayleigh lines and luminescence
bands are in separate datasets for the sample and blank; therefore there are four
curves used for analysis. Both methods are demonstrated below.
a
In the FluorEssence™ toolbar, click the
Quantum Yield
button
.
The
HJY_PLQY
window opens.
b
Browse with the
Browse
button
to find the
Sphere Correction
file.
The
Open
window appears:
Rayleigh peak
Luminescence