79
10.3 Default Settings
10.3.1 Meter Default Settings
Item
Selection item/Setting range
Default
values
Security
Security management
function
Enable, Disable
Disable
Hold condition Hold setting mode
EXACT, NORMAL, BRIEF,
TIME, CUSTOM, OFF (Manual)
NORMAL
In selecting
"TIME"
Time setting value
2 seconds to 999 seconds
10 seconds
In selecting
"CUSTOM"
Time setting value
2 seconds to 60 seconds
10 seconds
Conductivity variation
width
1 digit to 100 digits
1 digit
Salinity variation width
0.10 PPT to 10.00 PPT
0.30 PPT
Resistivity variation width 1 digit to 100 digits
1 digit
TDS variation width
0.1 mg/L to 100.0 mg/L
100.0 mg/L
Interval
memory
Interval memory function
Enable, Disable
Disable
Time setting value
1 second to 999 seconds
30 seconds
Содержание DS-72G
Страница 1: ...Instruction Manual COND METER DS 72G...
Страница 2: ......
Страница 12: ...For Your Safety X JPN EN61326 1 EN61010 1 II 2...
Страница 20: ......
Страница 104: ......
Страница 105: ......
Страница 106: ......