39
3.2 COND Measurement Setting
This section describes the procedures to set the conditions of COND measurement.
3.2.1 Cell Constant Setting
1.
Tap the channel setting and the measurement
item in the MEAS screen to set "CH2" and
"COND".
2.
Tap
and tap "CH2 MEAS SET".
3.
COND measurement setting items are
displayed.
You will see the remaining items by dragging.
4.
Select items and set the conditions.
The setting procedures for each item are
explained below.
1.
Tap the channel setting and the measurement
item in the MEAS screen to set "CH2" and
"COND".
2.
Press the CAL key.
3.
Tap
and tap "CH2 CAL SET".
A screen to select Cell Constant and Auto
Calibration is displayed.
4.
Tap
on the right of the Cell Constant item.
The cell constant setting screen is displayed.
Содержание DS-72G
Страница 1: ...Instruction Manual COND METER DS 72G...
Страница 2: ......
Страница 12: ...For Your Safety X JPN EN61326 1 EN61010 1 II 2...
Страница 20: ......
Страница 104: ......
Страница 105: ......
Страница 106: ......