![Horiba Scientific DS-72G Скачать руководство пользователя страница 100](http://html1.mh-extra.com/html/horiba-scientific/ds-72g/ds-72g_instruction-manual_2162306100.webp)
80
10.3.2 Measurement Condition Default Settings
(Can Be Set per Operator)
Item
Selection item/
Setting range
Default
values
Conductivity
measurement
condition
Alarm
condition
Upper limit
value setting
Enable, Disable
Disable
Lower limit
value setting
Enable, Disable
Disable
Upper limit
value
0.003
S/cm to 1.999 S/cm
1.999 S/cm
Lower limit
value
0.003
S/cm to 1.999 S/cm
0.003
S/cm
Measurement value unit
S/m, S/cm, FIX
S/cm
Temperature
setting
Temperature
setting
ATC (Automatic temperature
compensation),
MTC (Manual temperature
compensation)
ATC
Temperature
input value in
selecting "MTC"
0.0
C to 100.0
C
25.0
C
Temperature
conversion
Temperature
conversion
function
Pure water, Natural water,
Manual, Disable
Manual
Temperature
conversion
coefficient
0.00%/
C to 10.00%/
C
2.00%/
C
Reference
temperature
15.0
C to 30.0
C
25.0
C
Electrode
data
Model
None
lot No.
None
Salinity
measurement
condition
Alarm
condition
Upper limit
value setting
Enable, Disable
Disable
Lower limit
value setting
Enable, Disable
Disable
Upper limit
value
0.00 PPT to 80.00 PPT
(0.000% to 8.000%)
80.00 PPT
Lower limit
value
0.00 PPT to 80.00 PPT
(0.000% to 8.000%)
0.00 PPT
Measurement value unit
PPT, %
PPT
SAL type
NaCl, SEA water