10 EMC information
Guidance and Declaration - Electromagnetic Emissions
The device is suitable for use in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used in
such an environment.
Emission tests
Radio frequency
(RF) emissions
CISPR 11
Radio frequency
(RF) emissions
CISPR 11
Harmonic
emissions
IEC 61000-3-2
Voltage
fluctuations/flicker
emissions
61000-3-3
Group 1
Class A
Not applicable
Compliance
Electromagnetic environment - guidance
The 594810 uses RF energy only
for its internal function. Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment.
The 594810 is suitable for use in
all establishments other than domestic and
those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic
purposes.
Not applicable
The device meets the requirements of IEC 60601-1-2:2007.
PR
Display Range
Resolution
25 to 300 bpm
1 bpm
Accuracy
25 to 250bpm:±3 bpm (measured without motion)
Refreshing rate
1 s
Width × Height × Depth 56×124×30 mm
Max. weight
about 300g (full configuration, including the batteries)
Physical specifications