Select
[SET]
→
[ALn]
, and press the
Up/Down
button, to turn on/off the
alarm tone
5.3.3.3 Switch ON/OFF alarm Volume
!
WARNING
:
When the alarm sound is switched off, the pulse oximeter
will give no audible alarm tones even if a new alarm occurs. Therefore
the user should be very careful about whether to switch off the alarm
sound or not.
!
WARNING
:
Do not rely exclusively on the audible alarm system for
patient monitoring. Adjusting volume to a low level may result in a
hazard to the patient. Always keep the patient under close surveillance.
Press the
key for 1 second
,
the
key will light and the alarm
messages remain displayed. The alarm will stay silent for 60 seconds.
5.3.3.4
Alarm Silence Duration
!
WARNING
:
The 594810 will only keep the last setting. When the
594810 is turned on, user should check whether the current setting
meet conditions of patient.
5.3.4 When an Alarm Occurs
When an alarm occurs, observe the following steps to take proper actions:
1) Check the patient's condition.
2) Confirm the alarm parameter or alarm category.
3) Identify the source of the alarm.
4) Take proper action to eliminate the alarm condition.
Note:
The pulse rate alarm limit range is 25 to 250.The upper value of the
pulse rate low alarm limit is limited one number lower than the pulse rate
upper alarm limit. The pulse rate low alarm limit cannot be set equal to or
higher than the pulse rate upper alarm limit. In the same way, the pulse
rate upper alarm limit cannot be set equal to or lower than the pulse rate
low alarm limit.
Press button
[SET]
→
[SPO2 L]
and Press the Up/Down button, set the
SPO2 value between 20 and 100.
Note:
The %SpO2 alarm limit range is 20% to 100%. The upper value of
the %SpO2 low alarm limit is limited to the %SpO2 upper alarm limit. The
%SpO2 low alarm limit cannot be set equal to or higher than the %SpO2
upper alarm limit. In the same way,the lower value of the %SpO2 upper
alarmlimit is limited to the %SpO2 low alarm limit. The %SpO2 upper
alarm limit cannot be set equal to or lower than the %SpO2 low alarm limit.
5.3.3.2 Set SPO2 Low alarm Limit
-14-