![Hopkins 594810 Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/hopkins/594810/594810_instructions-for-use-manual_2161899025.webp)
Pulse Oximeter Probe Sensor Information
Model
Name
Testing
position
Intended
patient
Reusable Adult Finger Clip
SpO2 Sensor
Finger
Adult
(≥88lb)
594403
Adult Finger Sensor Usager
1) Hold the sensor with its opening towards the patient's index or
ring finger . The sensor should be oriented in such a way that
the sensor side with a fingertip sign is positioned on the top.
2) Insert the patient's index or ring finger into the sensor until the
fingernail tip rests against the stop at the end of the sensor. Adjust
the finger to be placed evenly on the middle base of the sensor.
Direct the cable along the top of the patient's hand.
3) Inspect the sensor site every 2 hours for skin integrity and relocate
the sensor every 2 hours according to steps 1-2 .
Reusable Neonatal Wrap
SpO2 Sensor
Foot,
hand or
finger
Child or
Neonatal/Infant
(Weight range)
594301
Neonatal Y-Probe Wrap Sensor Usage
1) Attach the sensor to a well perfused fleshy portion of the infant's
hand or foot.
2) Place the sensor to the patient with the light source (LED) on the
top of the hand or foot.
3) Secure the sensor with the Hopkins® Adjustable Y-Probe Sensor
Latex Free Belt and tighten gently. (Recommended) .
4) Apply a Hopkins® Y-Probe Wrap (#592012) on top to prevent
ambient light interference. A secure wrap with the probe sensor
centered in the middle of it will minimize any effects from
ambient light sources. Other wraps can be used to secure lead
wires to minimize the movement of the sensor itself.
(Recommended)
-25-