Measuring intensity
In order to achieve a consistent irradiation result, the intensity of the emitter
should be measured from time to time. The Hönle UV-Meter with surface
sensor, which is available as an accessory, is recommended for this task.
Fig. 8: Connecting the UV-Meter to the LED Cube 100
Measuring the UV irradiation intensity:
1. Remove the blind plug from the bottom right-hand corner of the door
2. Feed the UV-Meter cable through the opening in the housing
3. Place the surface sensor in the irradiation chamber
Fig. 9: Routing the UV-Meter cable to the LED Cube 100
4. Close the door
5. Switch on the UV-Meter (see UV-Meter Operating Instructions)
6. Start the irradiation procedure (see LED powerdrive Operating
Instructions)