12 

 

During the display of the danger 
alarm level, the LED visual 
alarm will be flashed twice and 
the audible danger alarm will be 
sounded twice.  The frequency 
of the audible danger alarm is 
higher than the frequency of the 
audible warning alarm.  The custom alarm if enabled is 
displayed similar to the danger alarm, but the LCD has 
the triangular warning symbol without the Danger or 
Warning symbol.  

 

 

 

 

For ToxiPro instruments with a toxic sensor and with the 
STEL and TWA alarms enabled, the ToxiPro will briefly 
show the STEL and TWA alarm levels. 

 

 

 

If the calibration due reminder is enabled and calibration 
is due, the following screens will be shown. 

 

 

 

Press the MODE button to acknowledge the calibration 
due reminder.  If the instrument shuts off when the 
MODE button is pressed with “cal due shown”, then the 
ToxiPro is configured with calibration due lockout 
enabled. 

See section 2.1.1 below for more details on the 
calibration due lockout function. 

The Calibration Due Reminder and Lockout functions 
may be enabled or disabled with BioTrak Software. 

The current gas readings screen will then be shown.  
Toxic sensor models should show 0 if the instrument is 
located in fresh air.  Oxygen sensor models should show 
20.9 in fresh air. 

In normal operation the screen alternates between gas 
reading and sensor type. 

 

2.1.1  Calibration due lockout 

When the ToxiPro’s calibration due lockout function is 
enabled and calibration is due, the calibration due 
warning will be shown at instrument start up and can not 
be bypassed.  The instrument must be calibrated 
immediately before it will show any gas readings.  The 
calibration may be performed manually, or by placing it 
in an IQ Express Dock.  Upon successful calibration, the 
ToxiPro will proceed to the current gas readings screen. 

The calibration due lockout function can be enabled or 
disabled with BioTrak software. 

2.2 

Backlight 

The ToxiPro includes a backlight that is automatically 
turned on during an alarm condition.  The backlight can 
also be manually activated while the current gas reading 
is displayed by pressing the MODE button once.  When 
manually activated, the backlight will automatically turn 
itself off in about twenty seconds.  If activated by an 
alarm condition, the backlight will remain on until the 
instrument is no longer in alarm. 

2.3 

Functions 

From the current gas reading 
screen, press the MODE button 
once to activate the backlight.  
Press MODE again to scroll 
through the screen options.   

Once the backlight has been lit, 
press MODE once more to view 
the MAX gas values screen.  The MAX figure represents 
the highest gas value reading that has been recorded by 
the instrument during the current operating session.  
ToxiPro O

2

 models will sequentially display both the 

highest and the lowest readings that have been recorded 
by the instrument in the current operating session. 

From the MAX screen, press the MODE button to view 
the time screen.  A dot below the colon between the 
hours and the minutes digits indicates afternoon/evening 
hours (pm). 

         

 

10:45 am                    10:45 pm 

If the STEL (Short Term Exposure Limit) alarm is 
enabled, press the MODE button 
once to view the STEL reading.  
The STEL value displayed 
represents the average value of 
the instrument readings for the 
target gas for the most recently 
completed 15 minutes of 
operation. 

If the TWA (Time Weighted Average) alarm is enabled, 
press the MODE button once 
more to view the current TWA 
value.  TWA values are 

Содержание ToxiPro CO

Страница 1: ...Reference Manual ToxiPro Single Gas Detector Part 13 264 N3 Version 4 ...

Страница 2: ...TIAL THAT THE INSTRUCTIONS IN THIS REFERENCE MANUAL BE READ FULLY UNDERSTOOD AND FOLLOWED ToxiPro NGSG Reference Manual Honeywell Analytics Part Number 13 264 N3 Version 4 Copyright 2015 by Honeywell Analytics Inc Lincolnshire Illinois 60069 All rights reserved No page or part of this operation manual may be reproduced in any form without written permission of the copyright owner shown above Honey...

Страница 3: ... Missing sensor during startup 9 1 3 6 Corrupt memory 10 1 3 7 Vibrating Alarm Error 10 1 3 8 Heartbeat Feature 10 1 4 Functions 10 1 4 1 Security beep 10 1 4 2 Latching alarms 10 1 4 3 Enhanced Latching Alarm 10 1 4 4 Silence warning alarms 10 1 5 IQ Dock Compatibility 10 1 6 Design components 10 1 7 Standard accessories 11 1 8 Value pack kits 11 2 FIELD OPERATION 11 2 1 Turning the ToxiPro on 11...

Страница 4: ...8 3 6 1 2 Forced fresh air zero calibration 18 3 6 2 Span calibration failure 18 3 6 2 1 Sensor out of range no CAL 18 3 6 2 2 Causes for span cal failure 18 3 7 Fresh air zero calibration in a contaminated atmosphere 18 4 MAINTENANCE 19 4 1 Replacing batteries 19 4 2 Replacing sensors 20 4 3 Proper Cleaning 20 4 4 Storage 20 5 PC INSTRUMENT COMMUNICATIONS 20 5 1 Event logging 20 5 2 Instrument fi...

Страница 5: ...vision 1 Groups E F G CSA Class I Division 1 Groups A B C D Temp Code T4 CSA Class II Division 1 Groups E F G UL International DEMKO A S 09 ATEX 0823124 II 1 G Ex ia IIC T4 IECEx UL 09 0011 Ex ia IIC T4 Ga KTL 13 KB4BO 0147X http www honeywellanalytics com media honeywell analytics products toxipro documents kcs certificate juarez toxipro 13kb4bo0147x pdf la en The ToxiPro NGSG is in conformity wi...

Страница 6: ...otentially dangerous readings The ToxiPro O2 should be periodically calibrated in fresh air A sensor that cannot be calibrated or is found to be out of tolerance must be replaced immediately An instrument equipped with a toxic gas sensor that fails calibration may not be used until testing with known concentration test gas determines that accuracy has been restored and the instrument is once again...

Страница 7: ...nt à l air libre Tout capteur ne pouvant être étalonné ou hors de sa plage de tolérance doit être remplacé immédiatement Si l étalonnage d un capteur de gaz toxiques n est pas concluant le détecteur concerné ne doit en aucun cas être utilisé Pour que celui ci soit de nouveau utilisable il convient préalablement de le tester en appliquant une concentration connue de gaz de test pour vérifier la pré...

Страница 8: ...8 ...

Страница 9: ...w In the danger state the LED s will flash red Two oxygen alarm set points have been provided for ToxiPro O2 instruments The danger alarm is sounded for oxygen deficiency and the warning alarm is sounded for oxygen enrichment ToxiPro instruments equipped with a toxic gas sensor have up to four alarm set points Warning Danger Custom TWA Time Weighted Average if enabled and STEL Short Term Exposure ...

Страница 10: ...atched When the alarm is latched and if the gas level goes below the alarm level the audible and visual alarms will continued If alarm latching is enabled and the instrument is in alarm for less than the latching duration the alarm will not latch when the gas level drops below the alarm level Latch Acknowledgement A latched alarm latch can be acknowledged by two methods The first is either by mode...

Страница 11: ...al the ToxiPro s display will be blank The ToxiPro O2 will show the negated horn icon or To initialize the instrument press the MODE button for 5 seconds ToxiPro O2 models will proceed with a 15 minute countdown while the oxygen sensor stabilizes Note The 15 minute sensor warm up period is only necessary for the initial start up of ToxiPro O2 models When the ToxiPro O2 warm up period concludes the...

Страница 12: ...bration may be performed manually or by placing it in an IQ Express Dock Upon successful calibration the ToxiPro will proceed to the current gas readings screen The calibration due lockout function can be enabled or disabled with BioTrak software 2 2 Backlight The ToxiPro includes a backlight that is automatically turned on during an alarm condition The backlight can also be manually activated whi...

Страница 13: ...mp test the calibration bottle icon will be displayed at the bottom of the current gas readings screen without the triangular warning symbol Note If an IQ Express Dock is unavailable a span calibration will also reset the bump test due notice 2 5 Low battery alarms When there are less than 7 days of battery life remaining the low battery icon will be lit When there are less than 8 hours of battery...

Страница 14: ...aspirated sample draw kit that is available separately During remote sampling the gas sample is drawn into the sensor compartment through the probe assembly and a length of tubing The ToxiPro and manual sample draw kit are delivered with polyester urethane fuel resistant tubing part number 53 001 This material is completely compatible with the toxic gases CO and H2S When using the ToxiPro and samp...

Страница 15: ...the calibration gas cylinder If readings are accurate there is no need to adjust your gas detector See section 3 4 for further details concerning the functional bump test If the readings are inaccurate the instrument must be span calibrated before further use as discussed in section 3 5 The accuracy of ToxiPro instruments equipped with toxic gas sensors should be checked periodically with known co...

Страница 16: ... lead to inaccurate and potentially dangerous readings 3 4 Functional bump testing toxic sensor versions The accuracy of ToxiPro instruments may be verified at any time by a simple functional bump test To perform a functional bump test do the following 1 Turn the ToxiPro on and wait at least three minutes to allow the readings to fully stabilize If the sensor has just been replaced it must be allo...

Страница 17: ...completes the span calibration it will emit three short beeps and display the maximum span calibration adjustment value for two seconds Note The maximum span calibration adjustment value shown is an indication of the relative health of the sensor As a sensor loses sensitivity the maximum adjustment level will approach the calibration gas concentration letting you know when the sensor is losing sen...

Страница 18: ...calibration will fail and the no and CAL screens will be alternately displayed After displaying no and CAL times the instrument will return to the current gas readings screen and the warning symbol and the calibration bottle icon will be shown which signifies that the instrument failed the last attempt to span calibrate Note If the Calibration Due Lockout is enabled the ToxiPro will turn itself of...

Страница 19: ... and the ToxiPro will show F 0 on the display and will beep 6 times upon startup It will then go into the set time date mode The next five steps describe programming the time and date 7 Press and release the MODE button to advance the setting by one Hold the MODE button down to scroll rapidly Afternoon and evening hours are differentiated by a dot below the colon on the display 10 a m 10 p m Once ...

Страница 20: ...he collected data to a PC See section 5 3 for instructions on initiating communications Warning Danger Custom STEL and TWA alarms are logged as events The event logger in ToxiPro version 9 20 or greater logs will nest events The word nested in this case refers to the logging of multiple events simultaneously When the instrument goes from Warning to Danger alarm condition then both Warning and Dang...

Страница 21: ...cations To initiate communications hold the MODE button as if you are turning the instrument off Once OFF or on is shown continue to hold MODE until the IrdA screen is shown and the LED IrDA port turns on Once IrdA is shown release the MODE button Then align the infrared port at the top front of the ToxiPro with the PC s infrared port to proceed with the download The LED IrDA port will blink durin...

Страница 22: ...lfur Dioxide SO2 1 1 100 1 100 50 150 0 2 n d n d 0 Nitrogen Dioxide NO2 5 8 1 0 100 90 270 0 n d n d 0 n d n d Chlorine Cl2 specific 0 3 0 n d 12 100 20 0 0 0 0 0 0 Chlorine Dioxide ClO2 specific 0 25 0 n d n d 60 100 0 0 0 n d 0 0 Ammonia NH3 54 47 21 0 5 0 n d 0 0 n d 0 0 0 100 0 0 Phosphine PH3 0 5 25 20 n d 0 1 n d n d n d 1 0 5 Hydrogen Cyanide HCN 0 5 350 160 5 100 20 60 0 1 100 65 5 50 n d...

Страница 23: ...ytics recommends the following procedure to establish a safe and prudent accuracy check schedule for your Honeywell Analytics instruments 1 During a period of initial use of at least 10 days in the intended atmosphere check the sensor response daily to be sure there is nothing in the atmosphere that is poisoning the sensor s The period of initial use must be of sufficient duration to ensure that t...

Страница 24: ...DAMAGE OF ANY KIND CONNECTED TO THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY Instrument Accessory Warranty Periods Product s Warranty Period PHD6 2 years from date of purchase ToxiPro MultiPro 2 years from date of purchase ToxiLtd 2 years after activation or 2 years after the Must Be Activated By date whichever comes first Battery packs and chargers sampling p...

Страница 25: ...Honeywell Analytics Inc 405 Barclay Blvd Lincolnshire IL 60069 800 711 6776 Fax 800 995 4992 www honeywellanalytics com ...

Отзывы: