14 

 

2.7 

Turning the ToxiPro off 

To turn the ToxiPro off, press 
and hold the MODE button down 
until the instrument chirps three 
times and OFF is displayed. 

Once OFF is displayed, release 
the MODE button.  The 
instrument has been successfully turned off when the 
display goes blank. 

If Always On Mode is enabled see section 2.8.1 for 
instructions for turning off the ToxiPro. 

2.8 

Always On Mode 

The ToxiPro may be configured 
so that it may not be turned off 
with the MODE button.  In 
Always On Mode, the MODE 
button is used to reset the MAX, 
STEL and TWA calculations.  
Press and hold the MODE button until the instrument 
chirps three times and “on” is displayed. 
After the “on” screen, the ToxiPro will proceed through 
the standard startup sequence until the current gas 
readings screen is shown.  MAX, STEL and TWA 
calculations will be reset.  

Always On Mode may be enabled or disabled with 
BioTrak software.  

Note:  The ToxiPro must be turned off to replace the 
battery.  See section 2.8.1. 

2.8.1  Turning Off in Always On Mode 

BioTrak software must be used to turn the ToxiPro 
off when it is in Always on Mode.  BioTrak is 
available at: 

http://www.honeywellanalytics.com

 

1. 

Open BioTrak and select ToxiPro.   

2. 

Select “Configure” in the software.   

3. 

Hold the MODE button down for about 10 
seconds until IrDA is shown and establish the 
connection with the PC. 

4. 

The Never Turn Off option appears at the bottom 
left of every page in the configuration window.  
Click the Turn Off button. 

2.9 

Sampling 

The ToxiPro may be used in either diffusion or sample-
draw mode.  In either mode, the gas sample must enter 
the sensor compartment for the instrument to register a 
gas reading. 

In diffusion mode, the atmosphere reaches the sensor 
by diffusing through the sensor port on the front of the 
instrument.  Normal air movements are enough to carry 
the sample to the sensor.  The sensor reacts quickly to 
changes in the concentration of the gas being measured. 

It is also possible to use the ToxiPro to sample remote 
locations with the hand-aspirated sample-draw kit that is 
available separately.  During remote sampling, the gas 
sample is drawn into the sensor compartment through 
the probe assembly and a length of tubing.  

 The ToxiPro and manual 

sample draw kit are delivered with polyester 
urethane (fuel-resistant) tubing part number 53-001.  
This material is completely compatible with the toxic 
gases CO and H

2

S.  When using the ToxiPro and 

sample draw kit to sample with any of the gas types 
and tubing lengths listed in the chart below, FEP-
lined tubing (part number 53-036) should be used. 

Gas Type 

Tubing Length 

CL

2

, CLO

Up to 10 ft/3m Max. 

HCN 

Up to 100 ft/30m Max. 

SO

2

, NO

2

PH

3

, NH

> 10 ft/3m up to 100 
ft/30m Max. 

Standard polyester urethane (fuel-resistant) tubing 
(part number 53-001) can be used otherwise.  Use of 
other types of tubing, or beyond recommended 
length(s), may cause inaccurate and potentially 
dangerous readings that could result in serious 
injury or death. 

2.9.1  Sample draw kit usage 

1. 

Connect the shorter section of tubing from the 
squeeze bulb to the sample draw adapter.  Then 
connect the longer section of tubing from the 
squeeze bulb to the sample probe. 

2. 

Slide the sample draw adapter into the sensor 
port on the ToxiPro and secure it. 

3. 

Cover the end of the sample draw probe 
assembly with a finger, and squeeze the aspirator 
bulb.  If there are no leaks in the sample draw kit 
components, the bulb should stay deflated for a 
few seconds. 

4. 

Insert the end of the sample probe into the 
location to be sampled. 

5. 

Squeeze the aspirator bulb several times to draw 
the sample from the remote location to the sensor 
compartment.  Allow one squeeze of the bulb for 
every one foot of sampling hose for the sample to 
reach the sensors.  Continue to squeeze the bulb 
for an additional 45 seconds or until readings 
stabilize. 

6. 

Note the gas measurement readings. 

  Hand aspirated remote 

sampling only provides continuous gas readings for 
the area in which the probe is located when the bulb 
is being continuously squeezed.   

Note:  Each time a reading is desired, it is necessary 
to squeeze the bulb a sufficient number of times to 
bring a fresh sample to the sensor compartment and 
to continue to squeeze the bulb until readings 
stabilize. 

Содержание ToxiPro CO

Страница 1: ...Reference Manual ToxiPro Single Gas Detector Part 13 264 N3 Version 4 ...

Страница 2: ...TIAL THAT THE INSTRUCTIONS IN THIS REFERENCE MANUAL BE READ FULLY UNDERSTOOD AND FOLLOWED ToxiPro NGSG Reference Manual Honeywell Analytics Part Number 13 264 N3 Version 4 Copyright 2015 by Honeywell Analytics Inc Lincolnshire Illinois 60069 All rights reserved No page or part of this operation manual may be reproduced in any form without written permission of the copyright owner shown above Honey...

Страница 3: ... Missing sensor during startup 9 1 3 6 Corrupt memory 10 1 3 7 Vibrating Alarm Error 10 1 3 8 Heartbeat Feature 10 1 4 Functions 10 1 4 1 Security beep 10 1 4 2 Latching alarms 10 1 4 3 Enhanced Latching Alarm 10 1 4 4 Silence warning alarms 10 1 5 IQ Dock Compatibility 10 1 6 Design components 10 1 7 Standard accessories 11 1 8 Value pack kits 11 2 FIELD OPERATION 11 2 1 Turning the ToxiPro on 11...

Страница 4: ...8 3 6 1 2 Forced fresh air zero calibration 18 3 6 2 Span calibration failure 18 3 6 2 1 Sensor out of range no CAL 18 3 6 2 2 Causes for span cal failure 18 3 7 Fresh air zero calibration in a contaminated atmosphere 18 4 MAINTENANCE 19 4 1 Replacing batteries 19 4 2 Replacing sensors 20 4 3 Proper Cleaning 20 4 4 Storage 20 5 PC INSTRUMENT COMMUNICATIONS 20 5 1 Event logging 20 5 2 Instrument fi...

Страница 5: ...vision 1 Groups E F G CSA Class I Division 1 Groups A B C D Temp Code T4 CSA Class II Division 1 Groups E F G UL International DEMKO A S 09 ATEX 0823124 II 1 G Ex ia IIC T4 IECEx UL 09 0011 Ex ia IIC T4 Ga KTL 13 KB4BO 0147X http www honeywellanalytics com media honeywell analytics products toxipro documents kcs certificate juarez toxipro 13kb4bo0147x pdf la en The ToxiPro NGSG is in conformity wi...

Страница 6: ...otentially dangerous readings The ToxiPro O2 should be periodically calibrated in fresh air A sensor that cannot be calibrated or is found to be out of tolerance must be replaced immediately An instrument equipped with a toxic gas sensor that fails calibration may not be used until testing with known concentration test gas determines that accuracy has been restored and the instrument is once again...

Страница 7: ...nt à l air libre Tout capteur ne pouvant être étalonné ou hors de sa plage de tolérance doit être remplacé immédiatement Si l étalonnage d un capteur de gaz toxiques n est pas concluant le détecteur concerné ne doit en aucun cas être utilisé Pour que celui ci soit de nouveau utilisable il convient préalablement de le tester en appliquant une concentration connue de gaz de test pour vérifier la pré...

Страница 8: ...8 ...

Страница 9: ...w In the danger state the LED s will flash red Two oxygen alarm set points have been provided for ToxiPro O2 instruments The danger alarm is sounded for oxygen deficiency and the warning alarm is sounded for oxygen enrichment ToxiPro instruments equipped with a toxic gas sensor have up to four alarm set points Warning Danger Custom TWA Time Weighted Average if enabled and STEL Short Term Exposure ...

Страница 10: ...atched When the alarm is latched and if the gas level goes below the alarm level the audible and visual alarms will continued If alarm latching is enabled and the instrument is in alarm for less than the latching duration the alarm will not latch when the gas level drops below the alarm level Latch Acknowledgement A latched alarm latch can be acknowledged by two methods The first is either by mode...

Страница 11: ...al the ToxiPro s display will be blank The ToxiPro O2 will show the negated horn icon or To initialize the instrument press the MODE button for 5 seconds ToxiPro O2 models will proceed with a 15 minute countdown while the oxygen sensor stabilizes Note The 15 minute sensor warm up period is only necessary for the initial start up of ToxiPro O2 models When the ToxiPro O2 warm up period concludes the...

Страница 12: ...bration may be performed manually or by placing it in an IQ Express Dock Upon successful calibration the ToxiPro will proceed to the current gas readings screen The calibration due lockout function can be enabled or disabled with BioTrak software 2 2 Backlight The ToxiPro includes a backlight that is automatically turned on during an alarm condition The backlight can also be manually activated whi...

Страница 13: ...mp test the calibration bottle icon will be displayed at the bottom of the current gas readings screen without the triangular warning symbol Note If an IQ Express Dock is unavailable a span calibration will also reset the bump test due notice 2 5 Low battery alarms When there are less than 7 days of battery life remaining the low battery icon will be lit When there are less than 8 hours of battery...

Страница 14: ...aspirated sample draw kit that is available separately During remote sampling the gas sample is drawn into the sensor compartment through the probe assembly and a length of tubing The ToxiPro and manual sample draw kit are delivered with polyester urethane fuel resistant tubing part number 53 001 This material is completely compatible with the toxic gases CO and H2S When using the ToxiPro and samp...

Страница 15: ...the calibration gas cylinder If readings are accurate there is no need to adjust your gas detector See section 3 4 for further details concerning the functional bump test If the readings are inaccurate the instrument must be span calibrated before further use as discussed in section 3 5 The accuracy of ToxiPro instruments equipped with toxic gas sensors should be checked periodically with known co...

Страница 16: ... lead to inaccurate and potentially dangerous readings 3 4 Functional bump testing toxic sensor versions The accuracy of ToxiPro instruments may be verified at any time by a simple functional bump test To perform a functional bump test do the following 1 Turn the ToxiPro on and wait at least three minutes to allow the readings to fully stabilize If the sensor has just been replaced it must be allo...

Страница 17: ...completes the span calibration it will emit three short beeps and display the maximum span calibration adjustment value for two seconds Note The maximum span calibration adjustment value shown is an indication of the relative health of the sensor As a sensor loses sensitivity the maximum adjustment level will approach the calibration gas concentration letting you know when the sensor is losing sen...

Страница 18: ...calibration will fail and the no and CAL screens will be alternately displayed After displaying no and CAL times the instrument will return to the current gas readings screen and the warning symbol and the calibration bottle icon will be shown which signifies that the instrument failed the last attempt to span calibrate Note If the Calibration Due Lockout is enabled the ToxiPro will turn itself of...

Страница 19: ... and the ToxiPro will show F 0 on the display and will beep 6 times upon startup It will then go into the set time date mode The next five steps describe programming the time and date 7 Press and release the MODE button to advance the setting by one Hold the MODE button down to scroll rapidly Afternoon and evening hours are differentiated by a dot below the colon on the display 10 a m 10 p m Once ...

Страница 20: ...he collected data to a PC See section 5 3 for instructions on initiating communications Warning Danger Custom STEL and TWA alarms are logged as events The event logger in ToxiPro version 9 20 or greater logs will nest events The word nested in this case refers to the logging of multiple events simultaneously When the instrument goes from Warning to Danger alarm condition then both Warning and Dang...

Страница 21: ...cations To initiate communications hold the MODE button as if you are turning the instrument off Once OFF or on is shown continue to hold MODE until the IrdA screen is shown and the LED IrDA port turns on Once IrdA is shown release the MODE button Then align the infrared port at the top front of the ToxiPro with the PC s infrared port to proceed with the download The LED IrDA port will blink durin...

Страница 22: ...lfur Dioxide SO2 1 1 100 1 100 50 150 0 2 n d n d 0 Nitrogen Dioxide NO2 5 8 1 0 100 90 270 0 n d n d 0 n d n d Chlorine Cl2 specific 0 3 0 n d 12 100 20 0 0 0 0 0 0 Chlorine Dioxide ClO2 specific 0 25 0 n d n d 60 100 0 0 0 n d 0 0 Ammonia NH3 54 47 21 0 5 0 n d 0 0 n d 0 0 0 100 0 0 Phosphine PH3 0 5 25 20 n d 0 1 n d n d n d 1 0 5 Hydrogen Cyanide HCN 0 5 350 160 5 100 20 60 0 1 100 65 5 50 n d...

Страница 23: ...ytics recommends the following procedure to establish a safe and prudent accuracy check schedule for your Honeywell Analytics instruments 1 During a period of initial use of at least 10 days in the intended atmosphere check the sensor response daily to be sure there is nothing in the atmosphere that is poisoning the sensor s The period of initial use must be of sufficient duration to ensure that t...

Страница 24: ...DAMAGE OF ANY KIND CONNECTED TO THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY Instrument Accessory Warranty Periods Product s Warranty Period PHD6 2 years from date of purchase ToxiPro MultiPro 2 years from date of purchase ToxiLtd 2 years after activation or 2 years after the Must Be Activated By date whichever comes first Battery packs and chargers sampling p...

Страница 25: ...Honeywell Analytics Inc 405 Barclay Blvd Lincolnshire IL 60069 800 711 6776 Fax 800 995 4992 www honeywellanalytics com ...

Отзывы: