background image

 

23 

Appendix C  Calibration Frequency Recommendation 

 

One of the most common questions 
that we are asked at Honeywell 
Analytics is: 

“How often should I 

calibrate my gas detector?” 

Sensor Reliability and Accuracy 

Today’s sensors are designed to 
provide years of reliable service.  In 
fact, many sensors are designed so 
that with normal use they will only lose 
5% of their sensitivity per year or 10% 
over a two-year period.  Given this, it 
should be possible to use a sensor for 
up to two full years without significant 
loss of sensitivity. 

Verification of Accuracy 

With so many reasons why a sensor 
can lose sensitivity and given the fact 
that dependable sensors can be key to 
survival in a hazardous environment, 
frequent verification of sensor 
performance is paramount. 

There is only one sure way to verify 
that a sensor can respond to the gas 
for which it is designed. That is to 
expose it to a known concentration of 
target gas and compare the reading 
with the concentration of the gas.  This 
is referred to as a “bump” test. This test 
is very simple and takes only a few 
seconds to accomplish. 

The safest 

course of action is to do a “bump” 
test prior to each day’s use.

 It is not 

necessary to make a calibration 
adjustment if the readings fall between 
90%* and 120% of the expected value. 
As an example, if a CO sensor is 
checked using a gas concentration of 
50 PPM it is not necessary to perform a 
calibration unless the readings are 
either below 45 PPM or above 60 PPM. 

*The Canadian Standards 
Association (CSA) requires 
combustible gas sensors to undergo 
calibration when the displayed value 
during a bump test fails to fall 
between 100% and 120% of the 
expected value for the gas. 

Lengthening the Intervals between 
Verification of Accuracy 

We are often asked whether there are 
any circumstances in which the period 
between accuracy checks may be 
lengthened.  

Honeywell Analytics is not the only 
manufacturer to be asked this question! 

One of the professional organizations 
to which Honeywell Analytics belongs 
is the Industrial Safety Equipment 
Association (ISEA). The “Instrument 
Products” group of this organization 
has been very active in developing a 
protocol to clarify the minimum 
conditions under which the interval 
between accuracy checks may be 
lengthened. 

A number of leading gas detection 
equipment manufacturers have 
participated in the development of the 
ISEA guidelines concerning calibration 
frequency. Honeywell Analytics’s 
procedures closely follow these 
guidelines.  

If your operating procedures do not 
permit daily checking of the sensors, 
Honeywell Analytics recommends the 
following procedure to establish a safe 
and prudent accuracy check schedule 
for your Honeywell Analytics 
instruments: 

1. 

During a period of initial use of at 
least 10 days in the intended 
atmosphere, check the sensor 
response daily to be sure there is 
nothing in the atmosphere that is 
poisoning the sensor(s). The 
period of initial use must be of 
sufficient duration to ensure that 
the sensors are exposed to all 
conditions that might have an 
adverse effect on the sensors.  

2. 

If these tests demonstrate that it is 
not necessary to make 
adjustments, the time between 
checks may be lengthened. The 
interval between accuracy 
checking should not exceed 30 
days.  

3. 

When the interval has been 
extended the toxic and 
combustible gas sensors should 
be replaced immediately upon 
warranty expiration. This will 
minimize the risk of failure during 
the interval between sensor 
checks.  

4. 

The history of the instrument 
response between verifications 
should be kept. Any conditions, 
incidents, experiences, or 
exposure to contaminants that 
might have an adverse effect on 
the calibration state of the sensors 
should trigger immediate re-

verification of accuracy before 
further use.  

5. 

Any changes in the environment in 
which the instrument is being 
used, or changes in the work that 
is being performed, should trigger 
a resumption of daily checking.  

6. 

If there is any doubt at any time as 
to the accuracy of the sensors, 
verify the accuracy of the sensors 
by exposing them to known 
concentration test gas before 
further use.  

Gas detectors used for the detection of 
oxygen deficiencies, flammable gases 
and vapors, or toxic contaminants must 
be maintained and operated properly to 
do the job they were designed to do. 
Always follow the guidelines provided 
by the manufacturer for any gas 
detection equipment you use! 

If there is any doubt regarding your gas 
detector's accuracy, do an accuracy 
check! All it takes is a few moments to 
verify whether or not your instruments 
are safe to use.  

One Button Auto Calibration 

While it is only necessary to do a 
“bump” test to ensure that the sensors 
are working properly, all current 
Honeywell Analytics gas detectors offer 
a one-button auto calibration feature. 
This feature allows you to calibrate a 
Honeywell Analytics gas detector in 
about the same time as it takes to 
co

mplete a “bump” test. The use of 

automatic bump test and calibration 
stations can further simplify the tasks, 
while automatically maintaining 
records. 

Don't take a chance with your life.   
Verify accuracy frequently! 

Please read also Honeywell Analytics’s 
application note: 

AN20010808 “Use of 

‘equivalent’ calibration gas mixtures”

.  

This application note provides 
procedures to ensure safe calibration of 
LEL sensors that are subject to silicone 
poisoning. 

Содержание ToxiPro CO

Страница 1: ...Reference Manual ToxiPro Single Gas Detector Part 13 264 N3 Version 4 ...

Страница 2: ...TIAL THAT THE INSTRUCTIONS IN THIS REFERENCE MANUAL BE READ FULLY UNDERSTOOD AND FOLLOWED ToxiPro NGSG Reference Manual Honeywell Analytics Part Number 13 264 N3 Version 4 Copyright 2015 by Honeywell Analytics Inc Lincolnshire Illinois 60069 All rights reserved No page or part of this operation manual may be reproduced in any form without written permission of the copyright owner shown above Honey...

Страница 3: ... Missing sensor during startup 9 1 3 6 Corrupt memory 10 1 3 7 Vibrating Alarm Error 10 1 3 8 Heartbeat Feature 10 1 4 Functions 10 1 4 1 Security beep 10 1 4 2 Latching alarms 10 1 4 3 Enhanced Latching Alarm 10 1 4 4 Silence warning alarms 10 1 5 IQ Dock Compatibility 10 1 6 Design components 10 1 7 Standard accessories 11 1 8 Value pack kits 11 2 FIELD OPERATION 11 2 1 Turning the ToxiPro on 11...

Страница 4: ...8 3 6 1 2 Forced fresh air zero calibration 18 3 6 2 Span calibration failure 18 3 6 2 1 Sensor out of range no CAL 18 3 6 2 2 Causes for span cal failure 18 3 7 Fresh air zero calibration in a contaminated atmosphere 18 4 MAINTENANCE 19 4 1 Replacing batteries 19 4 2 Replacing sensors 20 4 3 Proper Cleaning 20 4 4 Storage 20 5 PC INSTRUMENT COMMUNICATIONS 20 5 1 Event logging 20 5 2 Instrument fi...

Страница 5: ...vision 1 Groups E F G CSA Class I Division 1 Groups A B C D Temp Code T4 CSA Class II Division 1 Groups E F G UL International DEMKO A S 09 ATEX 0823124 II 1 G Ex ia IIC T4 IECEx UL 09 0011 Ex ia IIC T4 Ga KTL 13 KB4BO 0147X http www honeywellanalytics com media honeywell analytics products toxipro documents kcs certificate juarez toxipro 13kb4bo0147x pdf la en The ToxiPro NGSG is in conformity wi...

Страница 6: ...otentially dangerous readings The ToxiPro O2 should be periodically calibrated in fresh air A sensor that cannot be calibrated or is found to be out of tolerance must be replaced immediately An instrument equipped with a toxic gas sensor that fails calibration may not be used until testing with known concentration test gas determines that accuracy has been restored and the instrument is once again...

Страница 7: ...nt à l air libre Tout capteur ne pouvant être étalonné ou hors de sa plage de tolérance doit être remplacé immédiatement Si l étalonnage d un capteur de gaz toxiques n est pas concluant le détecteur concerné ne doit en aucun cas être utilisé Pour que celui ci soit de nouveau utilisable il convient préalablement de le tester en appliquant une concentration connue de gaz de test pour vérifier la pré...

Страница 8: ...8 ...

Страница 9: ...w In the danger state the LED s will flash red Two oxygen alarm set points have been provided for ToxiPro O2 instruments The danger alarm is sounded for oxygen deficiency and the warning alarm is sounded for oxygen enrichment ToxiPro instruments equipped with a toxic gas sensor have up to four alarm set points Warning Danger Custom TWA Time Weighted Average if enabled and STEL Short Term Exposure ...

Страница 10: ...atched When the alarm is latched and if the gas level goes below the alarm level the audible and visual alarms will continued If alarm latching is enabled and the instrument is in alarm for less than the latching duration the alarm will not latch when the gas level drops below the alarm level Latch Acknowledgement A latched alarm latch can be acknowledged by two methods The first is either by mode...

Страница 11: ...al the ToxiPro s display will be blank The ToxiPro O2 will show the negated horn icon or To initialize the instrument press the MODE button for 5 seconds ToxiPro O2 models will proceed with a 15 minute countdown while the oxygen sensor stabilizes Note The 15 minute sensor warm up period is only necessary for the initial start up of ToxiPro O2 models When the ToxiPro O2 warm up period concludes the...

Страница 12: ...bration may be performed manually or by placing it in an IQ Express Dock Upon successful calibration the ToxiPro will proceed to the current gas readings screen The calibration due lockout function can be enabled or disabled with BioTrak software 2 2 Backlight The ToxiPro includes a backlight that is automatically turned on during an alarm condition The backlight can also be manually activated whi...

Страница 13: ...mp test the calibration bottle icon will be displayed at the bottom of the current gas readings screen without the triangular warning symbol Note If an IQ Express Dock is unavailable a span calibration will also reset the bump test due notice 2 5 Low battery alarms When there are less than 7 days of battery life remaining the low battery icon will be lit When there are less than 8 hours of battery...

Страница 14: ...aspirated sample draw kit that is available separately During remote sampling the gas sample is drawn into the sensor compartment through the probe assembly and a length of tubing The ToxiPro and manual sample draw kit are delivered with polyester urethane fuel resistant tubing part number 53 001 This material is completely compatible with the toxic gases CO and H2S When using the ToxiPro and samp...

Страница 15: ...the calibration gas cylinder If readings are accurate there is no need to adjust your gas detector See section 3 4 for further details concerning the functional bump test If the readings are inaccurate the instrument must be span calibrated before further use as discussed in section 3 5 The accuracy of ToxiPro instruments equipped with toxic gas sensors should be checked periodically with known co...

Страница 16: ... lead to inaccurate and potentially dangerous readings 3 4 Functional bump testing toxic sensor versions The accuracy of ToxiPro instruments may be verified at any time by a simple functional bump test To perform a functional bump test do the following 1 Turn the ToxiPro on and wait at least three minutes to allow the readings to fully stabilize If the sensor has just been replaced it must be allo...

Страница 17: ...completes the span calibration it will emit three short beeps and display the maximum span calibration adjustment value for two seconds Note The maximum span calibration adjustment value shown is an indication of the relative health of the sensor As a sensor loses sensitivity the maximum adjustment level will approach the calibration gas concentration letting you know when the sensor is losing sen...

Страница 18: ...calibration will fail and the no and CAL screens will be alternately displayed After displaying no and CAL times the instrument will return to the current gas readings screen and the warning symbol and the calibration bottle icon will be shown which signifies that the instrument failed the last attempt to span calibrate Note If the Calibration Due Lockout is enabled the ToxiPro will turn itself of...

Страница 19: ... and the ToxiPro will show F 0 on the display and will beep 6 times upon startup It will then go into the set time date mode The next five steps describe programming the time and date 7 Press and release the MODE button to advance the setting by one Hold the MODE button down to scroll rapidly Afternoon and evening hours are differentiated by a dot below the colon on the display 10 a m 10 p m Once ...

Страница 20: ...he collected data to a PC See section 5 3 for instructions on initiating communications Warning Danger Custom STEL and TWA alarms are logged as events The event logger in ToxiPro version 9 20 or greater logs will nest events The word nested in this case refers to the logging of multiple events simultaneously When the instrument goes from Warning to Danger alarm condition then both Warning and Dang...

Страница 21: ...cations To initiate communications hold the MODE button as if you are turning the instrument off Once OFF or on is shown continue to hold MODE until the IrdA screen is shown and the LED IrDA port turns on Once IrdA is shown release the MODE button Then align the infrared port at the top front of the ToxiPro with the PC s infrared port to proceed with the download The LED IrDA port will blink durin...

Страница 22: ...lfur Dioxide SO2 1 1 100 1 100 50 150 0 2 n d n d 0 Nitrogen Dioxide NO2 5 8 1 0 100 90 270 0 n d n d 0 n d n d Chlorine Cl2 specific 0 3 0 n d 12 100 20 0 0 0 0 0 0 Chlorine Dioxide ClO2 specific 0 25 0 n d n d 60 100 0 0 0 n d 0 0 Ammonia NH3 54 47 21 0 5 0 n d 0 0 n d 0 0 0 100 0 0 Phosphine PH3 0 5 25 20 n d 0 1 n d n d n d 1 0 5 Hydrogen Cyanide HCN 0 5 350 160 5 100 20 60 0 1 100 65 5 50 n d...

Страница 23: ...ytics recommends the following procedure to establish a safe and prudent accuracy check schedule for your Honeywell Analytics instruments 1 During a period of initial use of at least 10 days in the intended atmosphere check the sensor response daily to be sure there is nothing in the atmosphere that is poisoning the sensor s The period of initial use must be of sufficient duration to ensure that t...

Страница 24: ...DAMAGE OF ANY KIND CONNECTED TO THE USE OF ITS PRODUCTS OR FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY Instrument Accessory Warranty Periods Product s Warranty Period PHD6 2 years from date of purchase ToxiPro MultiPro 2 years from date of purchase ToxiLtd 2 years after activation or 2 years after the Must Be Activated By date whichever comes first Battery packs and chargers sampling p...

Страница 25: ...Honeywell Analytics Inc 405 Barclay Blvd Lincolnshire IL 60069 800 711 6776 Fax 800 995 4992 www honeywellanalytics com ...

Отзывы: