6 - 10
If you have any issues with the calibration, check that the unit has been properly installed and verify that there is no rocking
motion when pressing on opposite corners of the scale. If you feel any movement, correct by adjusting the leveling screws
that support the scale on the check-stand. When they are correct, lock these down snugly and repeat all the calibration
steps.
Once you have completed the scale calibration, the scale must be verified by the appropriate country or state Weights and
Measures inspector. For details, contact your local government’s Weights and Measures department.
Calibration Verification
The following tests verify the scale's calibration. The following tests are based on a 2-digit accuracy setting for pounds and a 3-
digit accuracy setting for kilograms.
Increasing Load Test
1. Make sure there is no load on the scale platter and verify that the remote display reads:
0.00 lbs
or
0.000 kg
2. Place a 5 lb. or a 2.5kg weight on the center of the scale platter and verify the display reads:
5.00 lbs
or
2.5 kg
3. Place an additional 5 lb. or an additional 2.5kg weight on the center of the scale platter and verify the remote display
reads between:
9.99
and
10.01 lbs
or
4.995
and
5.005kg
4. Place an additional 10 lb. or 5kg weight on the center of the scale platter and verify the remote display reads between:
19.99
and
20.01 lbs
or
9.995
and
10.005kg
5. Place an additional 10 lb. or 5kg weight on the center of the scale platter and verify the remote display reads between:
29.99
and
30.01 lbs
or
14.995
and
15.005kg
Step 6.
Press the
F1
button to activate your selection.
The LCD display shows
SCALE CALIBRATE – WAIT SCALE RSP
for a few seconds,
then it changes to
Step 1 – Platter Empty, VOL BTN = Advance
. The top LED bar on
the scanner flashes once every few seconds.
Step 7.
Make sure platter is on correctly and is empty, then press the
Sound
button.
Step 8.
The LCD display leads you through the next step of placing a full weight of 30 lbs. on
the scale. The LEDs flash 3 times. Press the
Sound
button once the scale weight
settles on full weight.
Step 9.
Remove all the weight once the full capacity is stored. The LED’s flash 5 times.
Remove the weight, let the scale settle at zero lbs., then press the
Sound
button again
to accept 0 lbs.
Step 10.
The units stores the final weight, and beeps multiple times. The calibration process is
complete, and the scale restarts. Begin
Calibration Verification
, below.
Содержание Stratos 2700
Страница 1: ...Stratos 2700 Bioptic Scanner Scale User s Guide ...
Страница 6: ......
Страница 20: ...1 6 ...
Страница 36: ...2 16 ...
Страница 96: ...5 4 ...
Страница 110: ...6 14 ...
Страница 130: ...8 14 ...
Страница 203: ...9 73 Maximum Message Length ...
Страница 204: ...9 74 ...
Страница 212: ...11 2 Denmark 8 9 6 Norway 8 9 6 Spain 6 Country Codes ...
Страница 214: ...11 4 ...
Страница 218: ...12 4 ...
Страница 244: ...13 26 ...
Страница 250: ...14 6 ...
Страница 256: ...15 6 ...
Страница 270: ...B 2 ...
Страница 273: ...Programming Chart 0 1 2 3 4 5 6 7 8 ...
Страница 274: ...9 A B C D E F Save Discard ...
Страница 276: ......
Страница 277: ......
Страница 278: ... Honeywell Scanning Mobility 9680 Old Bailes Road Fort Mill SC 29707 www honeywellaidc com 2700 UG Rev D 3 14 ...