SPM Single Point Monitor
SPM Technical Handbook
4-12
Target Gas
Maximum Sample Line
Filter
Type A
Filter
Type B
Filter
Type C
Standard SPM
SPM with Extended
Sample Option
Hydrogen Sulfide
3 meters (10 ft)
30 meters (100 ft)
X
Hydrogen Sulfide-dry
X
Hydrogen Sulfide (low)
X
Nitrogen Dioxide
X
X
Ozone
no filter
Phosgene
X
Sulfur Dioxide**
X
X
Fluorine
3 meters (10 ft)
6 meters (20 ft)
X
X
Fluorine-Low Level
X
X
Boron Trifluoride
X
X
The Teflon-membrane particulate filter assembly, P/N 1830-0055 requires replacing this filter’s membrane, P/N
0235-1072 every 30 days.
* Hydrogen Bromide and Hydrogen Iodide are very water soluble. A high relative humidity (RH) will cause
sample loss due to absorption onto the sample lines. For these gases, use the table below as a guide to
determine the maximum sample length.
** Listing for current SO
2
- ppm calibration is the same for SO
2
- ppb calibration
Note:
When monitoring for gases using low-level calibrations, additional sample line restrictions may apply. Consult
Honeywell Analytics.
Caution
Acid scrubbing filters (P/N 710235) are not to be used with SPM sample lines.
Gas
RH of
Sample
Maximum Sample Length
(with extended sample
option)
Hydrogen Bromide
or
Hydrogen Iodide
<40%
100 ft. (30 m)
40-60%
20 ft. (6 m)
Содержание SPM
Страница 8: ...SPM Single Point Monitor Midas Technical Handbook 1 1 1 Operation ...
Страница 15: ...SPM Single Point Monitor SPM Technical Handbook 2 1 2 Installation ...
Страница 20: ...SPM Single Point Monitor Midas Technical Handbook 3 1 3 Maintenance ...
Страница 24: ...SPM Single Point Monitor Midas Technical Handbook 4 1 4 Specifications ...
Страница 36: ...SPM Single Point Monitor SPM Technical Handbook 5 1 5 Options ...
Страница 53: ...SPM Single Point Monitor SPM Technical Handbook A 1 A Bi Directional Communications Protocol ...
Страница 61: ...SPM Single Point Monitor SPM Technical Handbook B 1 B Warranty Statement ...