SPM Single Point Monitor
SPM Technical Handbook
4-10
Gas Name
TLV
(Threshold
Limit Value)
Default
Alarm
Level 1
1
Default
Alarm
Level 2
1
Sample
Time
(sec)
Chemcassette
®
Part Number
SP
EP
XPS
2
Oxidizers
Bromine (Br
2
)
100 ppb
50 ppb
100 ppb
60
711314
700313
Chlorine (Cl
2
)
0.5 ppm
0.25 ppm
0.5 ppm
30
1740-9308
Chlorine II (Cl
2
)
0.5 ppm
0.25 ppm
0.5 ppm
15
704007
Chlorine III (Cl
2
)
0.5 ppm
0.25 ppm
0.5 ppm
15
704308
700308
Chlorine (Cl
2
) Low Level
0.5 ppm
10 ppb
20 ppb
600
704007
1740-9308
Chlorine Dioxide (ClO
2
)
100 ppb
50 ppb
100 ppb
240
1750-9306
1740-9306
Fluorine (F
2
)
1 ppm
0.5 ppm
1 ppm
30
1750-9306
1740-9306
Fluorine (F
2
) Low Level
1 ppm
0.1 ppm
0.2 ppm
120
1750-9306
1740-9306
Hydrogen Peroxide (H
2
O
2
)
1 ppm
0.5 ppm
1 ppm
15
700278
700509
Nitrogen Dioxide (NO
2
)
3 ppm
1.5 ppm
3 ppm
30
703012
700305
Nitrogen Dioxide (NO
2
)
4
3 ppm
1.5 ppm
3 ppm
30
704308
700308
Nitrogen Dioxide (NO
2
)
5
3 ppm
1.5 ppm
3 ppm
240
1750-9306
1740-9306
Ozone (O
3
)
100 ppb
50 ppb
100 ppb
60
704514
Phosgene (COCl
2
)
100 ppb
50 ppb
100 ppb
30
702020
700307
1740-9307
Sulfur Dioxide (SO
2
)
2 ppm
1 ppm
2 ppm
15
705015
Sulfur Dioxide (SO
2
)
6
250 ppb
125 ppb
250 ppb
60
705027
700350
1
Other alarm levels can be made available - consult Honeywell Analytics
2
XPS Chemcassette
®
requires the appropriate calibration
3
Consult Honeywell Analytics for alarm levels
Caution
Gas concentrations below the full scale of the measuring
range will not be reported until after the sample time
duration has completed. For measured values above the
measuring range, an alarm will be issued immediately.
Notes:
“Sample time” in the table above represents the time required to:
• Detect < full-scale gas concentration
• Calculate and display concentration value
• Update the 4-20 mA signal
• Produce a RS-422 data packet (optional)
When concentration is > full-scale, response time is shortened in
proportion to the concentration.
Example:
Ammonia Full-scale = 75 ppm. Response time = 15 sec
Concentration
20 ppm
75 ppm
150 ppm
300 ppm
Response Time 15 seconds 15 seconds 7.5 seconds 3.8 seconds
Approximate Chemcassette
®
replacement intervals with no
target gas present:
SP Size:
2 weeks
EP Size:
30 days
XPS Size: 90 days
4
Nitrogen Dioxide using Chlorine III Chemcassette
®
5
Nitrogen Dioxide using Fluorine/Oxidizer Chemcassette
®
6
Earlier SO
2
calibration listed (ppm) uses the Sulfur Dioxide
Chemcassette (not Sulfur Dioxide low level Chemcassette)
Содержание SPM
Страница 8: ...SPM Single Point Monitor Midas Technical Handbook 1 1 1 Operation ...
Страница 15: ...SPM Single Point Monitor SPM Technical Handbook 2 1 2 Installation ...
Страница 20: ...SPM Single Point Monitor Midas Technical Handbook 3 1 3 Maintenance ...
Страница 24: ...SPM Single Point Monitor Midas Technical Handbook 4 1 4 Specifications ...
Страница 36: ...SPM Single Point Monitor SPM Technical Handbook 5 1 5 Options ...
Страница 53: ...SPM Single Point Monitor SPM Technical Handbook A 1 A Bi Directional Communications Protocol ...
Страница 61: ...SPM Single Point Monitor SPM Technical Handbook B 1 B Warranty Statement ...