Calibration
Step
Operation
Press
Display
4 Temperature
Calibration
TCal
Hold
(3 seconds)
tCAL
To cancel, press Mode button.
The display changes to a live temperature
reading, so you can continue to monitor the
sample.
or
to edit the Displayed Temperature value.
The displayed temperature value
flashes
at the current value and increments or
decrements
by 0. 1.
The output is held at its current percent of
range value.
Press and hold
or
to increment or
decrement by about 0.2/second
MODE
Momentarily
A new Temperature Offset value is
calculated.
If successful, the display will change to
online PV and the Output hold will
terminate.
If an error occurs, "FAIL" will display and
return to online PV. The previous
Temperature Calibration Offset value will
be retained. Refer to "Diagnostics" for error
messages.
20
DirectLine® DL423 Sensor Module User Manual
9/11