1.4 Specifications
Displayed pH
0-14 pH
Displayed Temperature Range
–10 °C to +110 °C (14 °F to 230 °F)
Process Temperature
–10 °C to +110 °C (14 °F to 230 °F)
Sensor Survivable Temperature Range
Durafet I
I & III
Meredian
II
–10 °C to +130 °C (14 °F to 266 °F)
0 °C to 110 °C (32 °F to 230 °F)
Electronics Module Ambient Temperature
–20 °C to +85 °C (–4 °F to +185 °F)
Output Type
4-20 mA (2-wire loop powered)
Output Scale
0-14 pH
Output Calibration
4-20 mA
Mating Connector Rating
Submersible to 6.1m (20’)
Output (Loop)
6m (19.7’) cordset or Shielded twisted pair with field
wiring connector
User Termination
Tinned leads
Cable Lengths
Sensor:
Durafet II & Durafet III:
6.1m (20’) or 15.24m (50’)
Meredian II:
3.65m (12’) or 6.1m (20’)
HPW7000:
.45m (1.5’) or 10.97m (36’)
Power
16-42 Vdc, 23mA max
Maximum load resistance:
250 ohms at 16 Vdc
600 ohms at 24 Vdc
1400 ohms at 42 Vdc
Local Display and Buttons
LCD 4-digit, 7-segment
Engineering Units
pH
degrees F
degrees C
Calibration Options
1 point Sample or 2 point Sample
Auto Buffer Recognition
Selections:
US, NIST, EURO
Solution Temperature Compensation
Selections:
0.00pH/10
°
C
–0.16pH/10
°
C
–0.32pH/10
°
C
Diagnostics
Sensor and electronics
Case
Weatherproof, corrosion-resistant plastic housing
, NEMA4X
Approvals
CE Mark for Industrial Applications
UL – General Purpose for Process Control
CSA – General Purpose
FM – CLI, DIV1, Groups A, B, C & D and CLI, Zone 0
AEx ia IIC (IS)
FM – CLI, DIV2, Groups A, B, C & D and CLI, Zone 2,
Groups IIC (N.I. Field Wiring)
Remote Mounting
Pipe, Wall, or DIN Rail
Dimensions
H 123 mm (4.84”) x W 48 mm (1.89”) x D 46 mm (1.81”)
Weight
Approximately 142 g (5.0 oz.)
Содержание DirectLine DL421
Страница 34: ...Spare Parts 28 DirectLine DL421 Sensor Module User Manual 2 04 ...
Страница 35: ...Spare Parts 2 04 DirectLine DL421 Sensor Module User Manual 29 ...
Страница 37: ...Appendix CE Mark Applications 2 04 DirectLine DL421 Sensor Module User Manual 31 ...
Страница 39: ...2 04 DirectLine DL421 Sensor Module User Manual 33 ...