Installation
2/04
DirectLine® DL421 Sensor Module User Manual
11
Figure 2-4 Remote Mounting
Cable
Connector
Housing
Electronics
Module
Electrode
Locking Screw in
Rear of Housing
Cable
Connector
Cover
Remote Wiring Cable
20 or 50 Feet
Remote Electronics
Module for
Durafet II or III Electrodes
1
2
3
4
5
6
Orange Reference Electrode Lead
White Temperature Compensation Lead
White Temperature Compensation Lead
Pigtail Shield Lead
1 Meg-ohm resistor (pre-installed by
Honeywell)
Coax Measuring Electrode Lead and
1 Meg-ohm resistor (pre-installed by
Honeywell)
Compression
Cap
Strain
Relief
Cover
Protective
Vinyl Boot
(remove)
Remote Electronics
Module for
Meredian II Electrodes
and HPW7000
Remote Wiring Cable
12 or 20 Feet for
Meredian II Electrodes,
1.5 or 36 Feet for
HPW7000
Electrode
Electronics
Module
1
2
3
6
5
4
1
2
3
4
5
6
Clear Reference Coax Lead
Black Temperature Sensor Lead
White Temperature Sensor Lead
Black/White Measuring Shield
Clear Measuring Coax Lead
No Connection
HPW7000 Electronics
Meredian II Electronics
Note: Three Green/White leads to DL421
mounting bracket screw. Mounting bracket
must be connected to earth ground.
1
2
3
6
5
4
Содержание DirectLine DL421
Страница 34: ...Spare Parts 28 DirectLine DL421 Sensor Module User Manual 2 04 ...
Страница 35: ...Spare Parts 2 04 DirectLine DL421 Sensor Module User Manual 29 ...
Страница 37: ...Appendix CE Mark Applications 2 04 DirectLine DL421 Sensor Module User Manual 31 ...
Страница 39: ...2 04 DirectLine DL421 Sensor Module User Manual 33 ...