7. Appendix: CE Mark Applications
In situations where the pH display appears to fluctuate (short deviations above 0.2 pH
points) due to field wiring electrical noise, the noise may be reduced by making the
additional ground connections as described in Figure 7-1.
Output cable
Qty. (2)
#8 screws
Connect yellow
wire to screw on
mounting bracket.
Carefully remove
shrink tubing cover
(Durafet II only)
Connect or mount
bracket to earth
ground.
Step 1
Step 2
Step 3
Remote cable option
Figure 7-1 Wiring for CE Mark Applications