FAXITRON CT USER
MANUAL
| HOLOGIC | REVISION 002
SECTION
Schedule of
Maintenance
8
8-11
6. Draw a line through the reconstructed wire like demonstrated below.
7. Once the line is drawn through the wire, from the menu select Analyze>Plot Profile (or
equivalently use keyboard shortcut Ctrl+K). A window will pop up showing an intensity
profile plot along the line drawn.
8. Select the window with the plotted line profile through the wire. It will look similar to the
example above.
9. On the line provided below, record the width of the profile at half the height of its peak.
Use the cursor to measure the X position on the left side of the width and on the right
side and the calculate the difference.
10. This value should be less than 0.4 mm for acceptance.
NOTE: If the 3D Resolution Test fails repeat the Geometric Calibration.
8.11.0 Cont.
Width
Содержание Faxitron CT
Страница 1: ...User s Manual from Hologic Document 5081 9544 Revision 002 Issued March 2020...
Страница 5: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 1 SECTION 1 Introduction...
Страница 14: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 2 SECTION 2 Installation and Set up...
Страница 20: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 SECTION Installation and Setup 2 2 7 180cm 70 70cm 27 65cm 25...
Страница 21: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 3 SECTION 3 Calibration...
Страница 26: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 4 SECTION 4 Quick Start Basic Operation of the Faxitron CT...
Страница 33: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 5 SECTION 5 Faxitron CT Software Operation...
Страница 61: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 6 SECTION 6 Compliance Requirements and Safety Measure...
Страница 63: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 7 SECTION 7 Service and Troubleshooting...
Страница 65: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 8 SECTION 8 Schedule of Maintenance...
Страница 78: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 9 SECTION 9 Revision History...