FAXITRON CT USER
MANUAL
| HOLOGIC | REVISION 002
SECTION
Installation
and Setup
2
2-6
X-RAY CMOS DETECTOR
Detector Size:
15cm x 15cm
Detector Pixels:
1548 x 1548 @ 99um pitch
Spatial Resolution:
100um or better
Frame Rate:
Typically, 2.5 fps for best image quality, (max 22 fps)
Saturation Dose
: 14uGy
Dynamic Range:
76dB
DQE (@0 lp/mm, RQA5):
70%
Scintillator:
Medical grade columnar cesium-iodide CsI(Tl)
EXTERIOR DIMENSIONS (Approximate)
cm
inches
Height
180
70
Width
70
27
Depth
65
25
COMPARTMENT INTERIOR DIMENSIONS (Approximate)
cm
inches
Height
19
8
Width
30
12
Depth
19
8
WEIGHT (Approximate)
kg
lbs
Net
140
315
Shipping
290
640
2.3.0 Cont.
Faxitron CT
Specifications
Содержание Faxitron CT
Страница 1: ...User s Manual from Hologic Document 5081 9544 Revision 002 Issued March 2020...
Страница 5: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 1 SECTION 1 Introduction...
Страница 14: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 2 SECTION 2 Installation and Set up...
Страница 20: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 SECTION Installation and Setup 2 2 7 180cm 70 70cm 27 65cm 25...
Страница 21: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 3 SECTION 3 Calibration...
Страница 26: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 4 SECTION 4 Quick Start Basic Operation of the Faxitron CT...
Страница 33: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 5 SECTION 5 Faxitron CT Software Operation...
Страница 61: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 6 SECTION 6 Compliance Requirements and Safety Measure...
Страница 63: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 7 SECTION 7 Service and Troubleshooting...
Страница 65: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 8 SECTION 8 Schedule of Maintenance...
Страница 78: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 9 SECTION 9 Revision History...