3-2
Calibration
To optimize image quality, the Faxitron CT Software and the hardware requires periodic
calibration. This calibration requires the acquisition of reference images.
On first installation the system should be allowed to reach the ambient room temperature and a
full calibration performed.
There are 3 calibrations for Faxitron CT which should be performed at each interval shown
Image Calibration
Daily
Geometric Calibration
Every 3 months
Uniformity Calibration
Every 3 months
Each day when the system is started up the user will be prompted to complete a warmup and
image calibration. In addition, it is recommended that the system be re-calibrated after the
system is moved from one room to another room with potentially different ambient
temperature/humidity settings.
If the system has been knocked or moved over an uneven floor, then all 3 calibrations should be
performed to maintain the image quality.
WARNING
The CT has a battery back up to move the unit without powering down, but the
system must be on mains power when capturing images, including calibration
.
Daily System Warmup and Image Calibration Sequence
Ideally, allow the system to warm up for 30 minutes before
calibration. This will ensure that the internal detector electronics
have stabilised, and the system will produce a higher quality image.
To calibrate the system at any time, click the Calibrate button on the
Startup Screen (main screen).
Then click Image Calibration.
3.1.0
Calibration
3.1.1
Image Calibration
Содержание Faxitron CT
Страница 1: ...User s Manual from Hologic Document 5081 9544 Revision 002 Issued March 2020...
Страница 5: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 1 SECTION 1 Introduction...
Страница 14: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 2 SECTION 2 Installation and Set up...
Страница 20: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 SECTION Installation and Setup 2 2 7 180cm 70 70cm 27 65cm 25...
Страница 21: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 3 SECTION 3 Calibration...
Страница 26: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 4 SECTION 4 Quick Start Basic Operation of the Faxitron CT...
Страница 33: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 5 SECTION 5 Faxitron CT Software Operation...
Страница 61: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 6 SECTION 6 Compliance Requirements and Safety Measure...
Страница 63: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 7 SECTION 7 Service and Troubleshooting...
Страница 65: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 8 SECTION 8 Schedule of Maintenance...
Страница 78: ...FAXITRON CT USER MANUAL HOLOGIC REVISION 002 9 SECTION 9 Revision History...