SECTION
Camera
Calibration
3
3-4
This time insert the white Geometric calibration phantom by placing it on the CT pedestal.
The phantom has a location stud that will center the phantom on the pedestal. Once positioned
click Okay.
Check the box to confirm the Phantom is placed and proceed by clicking Start.
The calibration will begin taking a full 360-degree image of the phantom which takes up to four
minutes.
3.1.2
Geometric
Calibration Cont.
Содержание Faxitron CT