10
11.
Red
“MRI”
Handpiece Receptacle - Attachment site for the MRI handpiece fitting with the
red
sleeve.
12.
Black Handpiece Receptacle
- Attachment site for the handpiece fitting with the black sleeve.
13.
Red
“US/STX”
Handpiece Receptacle - Attachment site for the US/STX handpiece fitting with the
red
sleeve.
14.
“Saline PV” (Pinch Valve)
- Attachment site for the silicone tubing section of the handpiece saline line. Controls the
flow of the saline to the handpiece.
ATEC Breast Biopsy and Excision System: Sapphire Unit Set Up
This section provides information on how to make all necessary connections to set up the ATEC Sapphire
unit.
The following connections will be explained:
1. ATEC Power Cord and Footswitch
2. ATEC Footswitch Connection
3. ATEC Vacuum Line Assembly
4. ATEC Handpiece or Eviva Biopsy Device Connection
NOTE:
Refer to Eviva IFU for Eviva biopsy device operating instructions and warnings.
ATEC Power Cord Connection and Circuit Breaker
CAUTION:
Grounding reliability can only be achieved when this equipment is connected to a receptacle marked “Hospital
Grade”. Check grounding continuity regularly.
WARNING:
The ATEC Sapphire
unit console was designed to be used with the ATEC Hospital Grade Power Cord
provided with the console. DO NOT use a different power cord with the ATEC Sapphire
unit. Using a different power cord
may create an electrical and fire hazard. DO NOT, under any circumstances, remove the ground wire or ground plug from
any power plug. DO NOT use an extension cord with this equipment. An adapter may be necessary depending on the outlet
used.
WARNING:
Ensure that the power cord is in good condition. A damaged power cord can be an electrical shock hazard.
When unplugging the unit, always grasp the plug at the insertion point and pull gently. NEVER pull on the cord to unplug the
unit.
Содержание ATEC SAPPHIRE
Страница 35: ...35...
Страница 36: ...36 Hologic Inc 250 Campus Drive Marlborough MA 01752 USA www hologic com 877 887 8767...