20
The following photos or diagrams offer specific replacement instructions for the external components of the console unit
referenced above.
FIELD REPLACEABLE ITEMS
Item
Instructions
Footswitch Assembly
Removal (with the console powered off)
1. Completely unwrap and remove footswitch cord from cord management plate
2. Disconnect upper and lower footswitch connectors by pulling out the top lock
collar and pushing in the bottom lock collar. Once disconnected, call Hologic
Technical Service for product retrun instructions.
Installing new footswitch assembly (with the console powered off)
1. Connect footswitch connectors so that the connectors point toward the back of
the console.
2. Wrap the footswitch counter clockwise around cord management plate.
Verification of Installation
1. Turn console power on.
2. The console will beep; and will then default to the biopsy mode.
3. Plug vacuum inlet line with thumb or finger.
4. With console in Biopsy mode, depress footswitch pedal and hold.
5. Check for solid
green
“Vacuum Ready” light, the console gives an audible beep
at the end of each cycle and cycling of the Pinch Valve.
6. Release footswitch pedal and verify that the solid
green
vacuum light turns off
and cycling stops. Vacuum releases from thumb.
Vacuum Line Assembly
Removal (with the console powered off)
1. Locate the white lock collar where the vacuum line assembly attaches to the
console.
2. Lift up on the vacuum line assembly approximately one-eighth inch, exposing a
gap between the white collar and black barb base.
3. Insert a small slotted screwdriver into the gap created, then press and hold down
the white collar.
4. While holding down the white collar, pull the vacuum line assembly up and out of
the vacuum line assembly connector.
Installation (with the console powered off)
1. Press the new vacuum line assembly into the vacuum line assembly connector.
2. Pull up to verify that it has locked into the connector. The vacuum line assembly
will have a small amount of vertical movement, but should not pull free of the
connector. Note: Do not twist line as you pull up.
Verification of Installation
1. Turn console power on.
2. You will hear the console beep; it will then default to the biopsy mode.
3. Plug the tip of the vacuum line assembly with thumb or finger.
4. Press the Setup button on the console.
5. Verify that the solid
green
vacuum ready light is illuminated, indicating that the
system is functioning properly.
6. Turn console power off.
Back of Console
Cord Management
Plate
FootSwtich Cord
FootSwtich
In-line hydrophobic filter
Vacuum line assembly
Black barb
Black barb base
White lock collar
Vacuum line
Connector