52
Device functions
PARAMETER SETTINGS
Settings range
Dependency
Settings steps
Accuracy
Press.
4 – 20 (30
1
) hPa
---
0.5 hPa
0.5 hPa
FLEX-Level
1 – 3
---
1 Level
---
P-Start
4 – 20 hPa
P-Min,
P-Max
0.5 hPa
0.5 hPa
P-Min
4 – 20 hPa
P-Start
0.5 hPa
0.5 hPa
P-Max
4 – 20 hPa
P-Start
0.5 hPa
0.5 hPa
Mask
Nose, Nose/Mouth
---
---
---
I-Pres.
E-Pres. – 20 (30
1
) hPa
E-Pres. – 20 (30
1
) hPa
0.5 hPa
0.5 hPa
E-Pres.
4 hPa – I-Pres.
4 hPa,
I-Pres.
0.5 hPa
0.5 hPa
I-Slope
0.0 – 1,0 s
I-Pres. - E-Pres.,
I:E
0.1 s
---
E-Slope
0.0 – 1.0 s
I-Pres. - E-Pres.
0.1 s
---
I-Trigger
1 – 3
---
1 Level
---
E-Trigger
1 – 3
---
1 Level
---
Frequency
5 – 30 bpm
---
1 bpm
1 bpm
I:E
0.4 – 8.0 s
I-Slope, E-Slope, Frequency
0.1 s
0.1 s
Delay Time
0 – 20 s
---
1 s
1 s
Backup freq
ON; OFF
---
---
---
VT min
1
OFF; 200 – 1500 ml
I-Pres. < 28 hPa
100 ml
100 ml
P-addit.
1
1 – 10 hPa
P-addit. + I-Pres.
30 hPa,
I-Pres. < 28 hPa,
VT min
200 ml
0.5 hPa
0.5 hPa
TI min
OFF; 0.1 s – (TI max - 0.1 s)
TI max - 0.1 s
0,1 s
0.1 s
TI max
OFF; 0.1 s – 4.0 s
TI min + 0.1 s
0.1 s
0.1 s
Ramp
0 – 60 min
---
1 min
1 min
P-Ramp
4 hPa – E-Pres.
APAP: 4 hPa – P-Start
CPAP/FLEX: 4 hPa – Press.
---
0.5 hPa
0.5 hPa
Mask Test
0 – 90 s
---
5 s
1 s
1
1 VECTOR et BILEVEL ST30