Device functions
49
Turbine
This menu item shows the total operating time of the turbine, the
fi
lter, and
the therapy time.
Turbine
The counter indicates the total turbine running time.
Filter
The counter indicates the operating hours of the
fi
lter since the last
reset.
Therapy
The counter indicates the operating time with the therapy pressure
applied.
Standby
This counter indicates the time the device was connected to the
power supply.
To display the counter,
fi
rst go to the Clinic-menu. Then select
Turbine
using the programing keys
and
. Press the enter key
. Each time
you press the enter key
, you see the next counter.
Backup freq
(only in mode S in VECTOR et BILEVEL, BILEVEL ST20
and ST30)
In the event of apnea lasting longer than 15 s, the patient will continue
to be ventilated with a frequency of 10 bpm and an I:E ratio of 1:2.
To activate or deactivate the Backup frequency,
fi
rst go to the Clinic-
menu. Then select
Backup
freq
using the programing keys
and
.
The PIN code is now requested. Once you have entered the PIN, you
can change the setting (“ON” or “OFF”) by pressing the enter key
.