Device functions
E-Slope (
only in modes S, T and ST)
Time of pressure decrease from I-Pressure to E-Pressure (pressure
decrease rate)
First set the I-Pressure, E-Pressure and the I-Slope. Then set the desired
time using the programing keys
and
. Press enter key
to con
fi
rm.
I-Pressure
I-Slope
E-Slope
0.3s
0.3s
E-Pressure
Trigger
(only in mode T)
To set the Trigger,
fi
rst go to the Clinic-menu. Then select
Trigger
using the programing keys
and
. Press the enter key
. Now the
device will prompt you to enter the PIN. Pressing the enter key
again
changes the setting from “OFF” to “ON” and vice versa.
I-Trigger
(only in modes S, T and ST)
The I-Trigger depends on the volume and determines the sensitivity for
detecting inspiration. It can be set to three levels.
1
10 ml
2
15 ml
3
25 ml
To set the I-Trigger,
fi
rst go to the Clinic-menu. Then select
I-Trigger
using the programing keys
and
. Press the enter key
. Now the device
will prompt you to enter the PIN. Now select the desired level by pressing
the keys
and
. Press enter key
to confirm. Next the parameter
E-Trigger is set.