1-3
Safety
FIGURE 1-3. AP3600 Single-Use Consumables
For routine maintenance, as discussed in this manual, use only Hitachi Chemical Diagnostics tubing in the AP3600 system to ensure
that there is no leakage.
Toxic Substances
The Wash Buffer concentrate contains sodium azide as a preservative. This substance is highly toxic by inhalation, ingestion, or skin
absorption. Sodium azide can react with lead or copper to form dangerously explosive metal azides in laboratory equipment and drain
traps. Use the following handling, storage, and disposal precautions when working with the Wash Buffer concentrate.
Handling
• Always use proper personal protective equipment, such as gloves and eye protection.
• Avoid sodium azide contact with lead or copper where there is potential for sodium azide to become metal azides.
• Avoid touching or replacing corroded materials if the corrosion may be due to sodium azide; contact the manufacturer for guidance on
how to clean or replace corroded parts.
Storage
• Do not store tubing or other materials that may have been in contact with solutions containing sodium azide.
• Collect sodium azide waste into a plastic reservoir and label the container with the proper name.
• Clearly label all medical devices, collection containers and waste disposal fixtures that use products containing sodium azide.
Pette with Pette Tip installed
Sample Cup
Transfer Tips
Содержание OPTIGEN AP3600
Страница 1: ...OPTIGEN AP3600 AUTOMATED INSTRUMENT USER MANUAL ...
Страница 2: ...OPTIGEN AP3600 Automated Instrument UserManual Rx Only ...
Страница 6: ...OPTIGEN AP3600 AUTOMATED INSTRUMENT USER MANUAL Glossary E 1 v Table of Contents ...
Страница 14: ...1 6 Safety This page intentionally left blank ...
Страница 42: ...OPTIGEN AP3600 AUTOMATED INSTRUMENT USER MANUAL This page intentionally left blank 6 2 Work List Screen ...
Страница 46: ...7 4 Results Screen OPTIGEN AP3600 AUTOMATED INSTRUMENT USER MANUAL This page intentionally left blank ...
Страница 54: ...8 8 Utilities Screen This page intentionally left blank ...
Страница 61: ...9 7 System Settings Screen This page intentionally left blank ...
Страница 62: ......
Страница 66: ...10 4 End of Run This page intentionally left blank ...
Страница 68: ...OPTIGEN AP3600 AUTOMATED INSTRUMENT USER MANUAL This page intentionally left blank 11 2 Troubleshooting ...
Страница 70: ...OPTIGEN AP3600 AUTOMATED INSTRUMENT USER MANUAL This page intentionally left blank A 2 Regulatory Compliance ...
Страница 82: ...D 4 Maintenance This page intentionally left blank ...
Страница 84: ...OPTIGEN AP3600 AUTOMATED INSTRUMENT USER MANUAL This page intentionally left blank E 2 Glossary ...