5-1
Set Up & Run Wizard
Set Up & Run Wizard
The Set Up & Run wizard is a sequence of software screens and instructions for starting the OPTIGEN assay. Patient sera should be
prepared before launching the wizard. For more information on preparing samples before placing them into the AP3600 instrument, refer
to Chapter 3, Sample Preparation.
Prepare AP3600 Instrument
Perform the following steps to prepare the AP3600 instrument.
1. Confirm the transfer tip rack is filled and placed in the rack holder. Refer to Figure 5-1.
Note:
The assay will not start if the rack is not fully populated.
FIGURE 5-1. AP3600 Transfer Tip Rack, Sample Vial Racks, and Sample Cup Carousel
2. Fill the sample cup carousel with new cups. Refer to Figure 5-1.
3. Fill the DI water and wash buffer bottles. Confirm there is sufficient Wash Buffer and DI water available to run the number of tests
planned.
4. Load the reagent bottles (Antibody, Photoreagent I, and Photoreagent II) on the right side of the AP3600 instrument at this time.
5. Empty the waste container. There are both high and low sensors present for the waste container to ensure that the container does not
overflow during testing. Refer to Figure 5-2.
Transfer
Tip
Rack
Sample Cup Carousel
Sample Vial
Racks
Содержание OPTIGEN AP3600
Страница 1: ...OPTIGEN AP3600 AUTOMATED INSTRUMENT USER MANUAL ...
Страница 2: ...OPTIGEN AP3600 Automated Instrument UserManual Rx Only ...
Страница 6: ...OPTIGEN AP3600 AUTOMATED INSTRUMENT USER MANUAL Glossary E 1 v Table of Contents ...
Страница 14: ...1 6 Safety This page intentionally left blank ...
Страница 42: ...OPTIGEN AP3600 AUTOMATED INSTRUMENT USER MANUAL This page intentionally left blank 6 2 Work List Screen ...
Страница 46: ...7 4 Results Screen OPTIGEN AP3600 AUTOMATED INSTRUMENT USER MANUAL This page intentionally left blank ...
Страница 54: ...8 8 Utilities Screen This page intentionally left blank ...
Страница 61: ...9 7 System Settings Screen This page intentionally left blank ...
Страница 62: ......
Страница 66: ...10 4 End of Run This page intentionally left blank ...
Страница 68: ...OPTIGEN AP3600 AUTOMATED INSTRUMENT USER MANUAL This page intentionally left blank 11 2 Troubleshooting ...
Страница 70: ...OPTIGEN AP3600 AUTOMATED INSTRUMENT USER MANUAL This page intentionally left blank A 2 Regulatory Compliance ...
Страница 82: ...D 4 Maintenance This page intentionally left blank ...
Страница 84: ...OPTIGEN AP3600 AUTOMATED INSTRUMENT USER MANUAL This page intentionally left blank E 2 Glossary ...