6.5 Input / Output Functions (Interface for External Control)
58
External I/O
Input/Output signals
Input
Signal types
Input method
Electrical specification
Trigger (TRIGGER), fixture capacitance open correction execution (OPEN_CX),
fixture resistance open correction execution (OPEN_IR),
contact check execution (C.CHECK)
Photocoupler-isolated input
LOW: 0.5 V or less
HIGH: 24 V±10%
Output
Signal types
Output method
Electrical specification
Measured value comparative judgment results (Hi1 to Hi8, IN1 to IN8, LO1 to LO8),
contact check judgment results (NO_CONTACT1 to NO_CONTACT8),
error (ALARM), measurement calculation complete (EOM),
analog measurement complete (INDEX)
Photocoupler-isolated output, open collector output
LOW: 0.5 V or less
HIGH: 5 to 24 V (depends on external power source voltage)
Output current: 5 mA max.
Connector
57RE-40500-730B (50-pin: DDK)