4.6 Message List
37
OST?
Returns the fixture capacitance open correction value (fixture
capacitance) as a response.
[Format]
d1 (operation specification)
0: Returns the capacitance without performing open correction.
1: Performs open correction and then returns the capacitance.
If an error occurs, this command will return the value 999.9.
[Response]
d1 (CH1 fixture capacitance: 0 to 99.9)
d2 (CH2 fixture capacitance: 0 to 99.9)
d3 (CH3 fixture capacitance: 0 to 99.9)
d4 (CH4 fixture capacitance: 0 to 99.9)
d5 (CH5 fixture capacitance: 0 to 99.9)
d6 (CH6 fixture capacitance: 0 to 99.9)
d7 (CH7 fixture capacitance: 0 to 99.9)
d8 (CH8 fixture capacitance: 0 to 99.9)
Error: 999.9
Note: Open correction must be performed once before a contact
check can be performed.
Note: Omitted parameters are treated as 0.
[Format]
OST? d1
d1: NR1 format
[Response] d1,d2,d3,d4,
d5,d6,d7,d8
d1: NR2 format
d2: NR2 format
d3: NR2 format
d4: NR2 format
d5: NR2 format
d6: NR2 format
d7: NR2 format
d8: NR2 format
Measured value comparison and judgment function
CMP
Comparative measurement mode setting
d1 (Execute comparison: 0 to 1)
0: OFF 1: ON
d1 (Mode: 0 to 2)
0: HI 1: IN 2: LO
d3 (Upper limit comparison value)
(-9.9999E+30 to 9.9999E+30)
d4 (Lower limit comparison value)
(-9.9999E+30 to 9.9999E+30)
Note: Always set parameters so that d3
≥
d4. Failure to do so will
cause the current settings to be applied.
Note: The d2, d3, and d4 parameters are valid even when compar-
ison execution is set to OFF.
(They will be saved as the current settings.)
[Format]
CMP d1,d2,d3,d4
d1: NR1 format
d2: NR1 format
d3: NR3 format
d4: NR3 format
CMP?
Comparative measurement mode query
The contents of responses are the same as the settings.
[Format]
CMP?
[Response] d1,d2,d3,d4
Fixture resistance open correction function
OCM
Fixture resistance open correction mode selection
d1 (Selection: 0 to 1)
0: OFF (Disables use of correction value in measured value
calculations.)
1: ON (Enables use of correction value in measured value cal-
culations.)
[Format]
OCM d1
d1: NR1 format
OCM?
Fixture resistance open correction mode query
The contents of responses are the same as the settings.
[Format]
OCM?
[Response] d1
OCL
Performs fixture resistance open correction once and saves the
correction value.
d1 (Channel specification)
1 to 255: Specifies the channel for which to perform correction as
the weight of bits 0 (channel 1) to 7 (channel 8).
[Format]
OCL d1
d1: NR1 format
OIR?
Fixture resistance open value query
The contents of responses are the same as the settings.
Note: The value for the current channel is returned as a query.
Note: The AD converted values for the instrument's internal
ammeter's seven ranges are used as the return values.
Note: A return value of 32768 indicates that correction was not
performed due to an error.
[Format]
OIR?
[Response] d1,d2,d3,d4,d5,
d6,d7
d1 to d7:
NR1 format
Command
Description
Formats
Содержание SM7810
Страница 19: ...2 5 Turning the Power On and Off 16 ...
Страница 51: ...5 3 Internal Circuitry 48 ...
Страница 65: ...7 4 Cleaning 62 ...