2.5 Turning the Power On and Off
15
Turning Power On
Turning Power Off
2.5 Turning the Power On and Off
Before turning the instrument on, make sure the supply voltage matches
that indicated on its power connector. Connection to an improper supply
voltage may damage the instrument and present an electrical hazard.
Rear Panel
POWER switch
Turn the POWER switch on ( ).
The power indicator and LCD screen on the front of the instrument will light up.
When the power is turned on, the same setting as when the power was last turned off appears
(backup function).
When powered up for the first time, the instrument will be configured with its default settings.
However, the LCD screen is not backed up.
Before Starting Measurement
To obtain precise measurements, provide about 60 minutes warm-up after turning power on.
Before turning the instrument off, turn off measurement power source output.
Turn the POWER switch off ( ).
If a power outage (e.g., breaker trip) occurs when the instrument is on, it will
automatically turn on again when power is restored.