7.4 MR8904 CAN Editor General Specifications
60
7.4
MR8904 CAN Editor General Specifications
Supported operating systems
Windows XP 32-bit
Windows Vista/7/8 32-bit and 64-bit
Recommended display
resolution
1280 x 960
Interface
USB
Supported measuring
instrument
Hioki MR8875 Memory HiCorder
CAN definition settings
1. Item number: Starting at 1
2. CAN message ID: 0 to 1FFFFFFF H
3. Start position: 0 to 63
4. Data length: 1 to 32
5. Data order: U/L (Motorola), L/U (Motorola), L/U (Intel)
6. Sign: Unsigned, 1-signed, 2-signed
7. Signal name: Up to 40 characters
8. Label name: Up to 16 characters
9. Scaling unit: Up to 7 characters
10. Scaling value: Bit rate/offset value or maximum value/minimum value
Definition file memo
Memos consisting of up to 50 characters can be stored in the CAN definition file.
Password
A password of up to 20 characters can required in order to edit CAN definitions.
CANdb file
1. Supported: Loading, conversion to CAN definition data file format (.CDF), regis-
tration to register list
2. Not supported: Editing
3. Data of 33 bits and more is not supported.
4. Data order: Conversion from Motorola (CANdb file) to U/L (Motorola)
5. Conversion from signed (CANdb file) to 2-signed; IEEE float and double
(CANdb file) are not supported.
6. Conversion from signal name (CANdb file) to label
7. Conversion from comment (CANdb file) to signal name
Register list settings
1. CAN input port setting: Select port 1 or port 2.
2. Item number: 1 to 200
3. Display upper and lower limit value settings on Memory HiCorder
Set unit position
Unit 1 to Unit 4
CAN communications settings
1. Set (2) to (5) for each port.
2. Interface: High-speed, low-speed, or single-wire
3. Terminator: On/off (on is valid with high-speed interfaces only)
4. ACK: On/off
5. Baud rate: Auto (valid during ACK off only)
High-speed CAN: 1M, 800k, 500k, 250k, 125k, 100k, 83.3k, 62.5k, 50 kbps
Low-speed CAN: 125k, 100k, 83.3k, 62.5k, 50k, 33.3k, 20k, 10kbps
Single-wire CAN: 83.3k, 62.5k, 50k, 33.3k, 20k, 10kbps
Analog channel settings
1. Number of channels: 15
2. Definitions on register list within 16 bits are allocated to 1 channel.
3. Definitions on register list from 17 to 32 bits are allocated to 2 channels.
Logic channel settings
1. Number of channels: 16
2. Definitions on register list within 16 bits and their bit position are allocated.
3. Definitions on register list are allocated as ID triggers.
Содержание MR8904
Страница 1: ...MR8904 Instruction Manual CAN UNIT EN Sept 2018 Revised edition 3 MR8904A981 03 18 09H...
Страница 2: ......
Страница 10: ...Usage Notes 6...
Страница 14: ...10...
Страница 24: ...3 3 Connecting the Unit to a PC 20...
Страница 66: ...7 4 MR8904 CAN Editor General Specifications 62...
Страница 69: ......
Страница 70: ......
Страница 71: ......
Страница 72: ......