Input, Output, and Measurement Specifications
Earth potential measurement
DC/AC auto-
detection range
0.3 V DC ±0.2 V or higher is detected as DC
Manual switching function provided
Pulsating flow with superimposed AC component that
periodically zero-crosses is detected as AC.
Measurement
method
Average value rectification, RMS value indication
Terminal
for voltage
measurement
3-pole method:
Between S terminal (+) and E terminal (−)
2-pole method:
Between H terminal (+) and E terminal (−)
Overload
protection
360 V AC (for 1 min., between each terminal)
360 V DC (for 1 min., between each terminal)
Display refresh
interval
Within 1 s
Input resistance
4 M
Ω
or higher (DC / 50 Hz / 60 Hz)
Response time
Within 2 s (when input voltage is changed from 0 V to
30 V)
Effect of
temperature
Accuracy × 1.0 ( −10°C to 50°C, applied in the range
excluding 18°C to 28°C)
Accuracy × 2.0 (−25°C to −10°C, 50°C to 65°C)
Range
configuration
Display range
30 V rms
Maximum display value 30.0 V rms
Resolution
0.1 V
Accuracy
DC
±1.3% rdg ±4 dgt
50 Hz/60 Hz
±2.3% rdg ±8 dgt
HIOKI FT6031B961-00
Содержание FT6031-50
Страница 2: ...HIOKI FT6031B961 00...
Страница 6: ...iv Contents HIOKI FT6031B961 00...
Страница 20: ...14 Usage Notes HIOKI FT6031B961 00...
Страница 25: ...19 Part Names and Functions Operation buttons 1 2 3 4 6 7 5 HIOKI FT6031B961 00...
Страница 90: ...84 Error Display HIOKI FT6031B961 00...
Страница 94: ...88 Measurement Principle HIOKI FT6031B961 00...
Страница 95: ...HIOKI FT6031B961 00...
Страница 96: ...HIOKI FT6031B961 00...
Страница 97: ...HIOKI FT6031B961 00...