50
Precise Measurement for Earth Resistance
(Precise Measurement Method, 3-pole Method)
Disabling the continuous measurement function
2
1
1
Press the
MEASURE
button
during continuous measurement.
Continuous measurement will stop.
2
Press and hold the
3P/2P
(
CONT
)
button for at least 1 s until the
indicator is no longer
shown on the display.
Continuous operating time
When using four AA alkaline batteries (LR6) (reference value at 23°C)
•
Approx. 3.5 hours (3-pole method, auxiliary earthing resistance of 100
Ω
,
measuring 10
Ω
using the 20
Ω
range,
without Z3210 installed)
• Approx. 3 hours (3-pole method,
auxiliary earthing resistance of 100
Ω
,
measuring 10
Ω
using the 20
Ω
range,
with Z3210 installed, using wireless communications)
HIOKI FT6031B961-00