Simplified Measurement for Earth Resistance
(Simplified Measurement Method, 2-pole Method)
Performing zero adjustment
The range that the instrument can provide zero adjustment for is
3
Ω
or less. If the measured value exceeds 3
Ω
, the message
Err
0
Ω
ADJ
appears and zero adjustment will not be executed.
1
3
4
6
2
7
H(C)
E
5
1
Remove the cover of the
measurement terminals.
2
Connect the E terminal and
H(C) terminal of the instrument
respectively with the L9787 Test
Lead (black) and L9787 Test Lead
(red).
3
Press the
Power
button to turn it
on.
4
Press the
3P/2P
button to display
(2-pole method).
A measured value (approx. 0 V) of the earth
potential appears.
5
Connect (short-circuit) the tips of
test leads.
6
Press the
MEASURE
button.
A measured value of approx. 0
Ω
appears
in approx. 3 s, and
lights up.
When the
indicator appears below
the measurement terminal, a test lead
may not have been connected to the
measurement terminal, or the lead may
have a break. Check the connection and
perform cable check with a tester etc.
7
Press the
0
Ω
ADJ
button.
lights up and then zero
adjustment is completed.
How to disable zero adjustment
Press and hold the
0
Ω
ADJ
button for at least 1 s while
is on.
HIOKI FT6031B961-00
Содержание FT6031-50
Страница 2: ...HIOKI FT6031B961 00...
Страница 6: ...iv Contents HIOKI FT6031B961 00...
Страница 20: ...14 Usage Notes HIOKI FT6031B961 00...
Страница 25: ...19 Part Names and Functions Operation buttons 1 2 3 4 6 7 5 HIOKI FT6031B961 00...
Страница 90: ...84 Error Display HIOKI FT6031B961 00...
Страница 94: ...88 Measurement Principle HIOKI FT6031B961 00...
Страница 95: ...HIOKI FT6031B961 00...
Страница 96: ...HIOKI FT6031B961 00...
Страница 97: ...HIOKI FT6031B961 00...