88
Table 7.3 Setting Control Program Messages 7/13
Mnemonic Description
Format
CMP
Comparison Measuring mode setting
d1 (Comparison execute: 0 ~ 1)
0: OFF 1: ON
d2 (Mode: 0 ~ 2)
0: HI 1: IN 2: LO
d3 (Upper-limit comparison value)
(-9.999E+30 ~ 9.999E+30)
d4 (Lower-limit comparison value)
(-9.999E+30 ~ 9.999E+30)
[Note]
1. Must be (d3 > d4).
If not complied with, currently set value will be
used.
2. Comparison execute: d2, d3 and d4 are
enabled even when OFF.
(Retained as current set value)
[Format]
CMP
d1: NR1 format
d2: NR1 format
d3: NR3 format
d4: NR3 format
CMP?
Query response of Comparison Measuring mode
is same as setting.
[Format]
CMP?
[Response]
d1, d2, d3, d4
WCP
Work capacity setting
Capacity of work for open correction value
calculation
d1 (0.5 ~ 99.9) pF
d1:CH1 d2:CH2 d3:CH3 d4:CH4
[Format]
WCP d1, d2, d3, d4
WCP?
Query response of work capacity is same as
setting.
[Format]
WCP?
[Response]
d1, d2, d3, d4
ACL
Auto self-calibration (current range calibration)
setting
d1 (Auto self-calibration done and not done: 0 ~
1)
0: OFF
1: ON
d2 (Auto self-calibration interval time)
(10 ~ 9999s)
[Format]
ACL d1, d2
d1: NR1 format
d2: NR1 format
ACL?
Query response of auto self-calibration (current
range calibration) is same as setting.
[Format]
ACL?
[Response]
d1, d2
Содержание DSM-8542
Страница 1: ...DIGITAL SUPER Instruction Manual DSM 8542 July 2013 Revised edition 1 DS8542A981 01 13 07H MEGOHMMETER...
Страница 2: ......
Страница 21: ...11 3 2 Block Diagram of DSM 8542...
Страница 89: ...79 Syntax Chart Fig 7 1 Message Syntax Chart...
Страница 90: ...80 Command program header Fig 7 2 Program Header Syntax Chart...
Страница 91: ...81 Fig 7 3 Syntax Chart of Data Part...
Страница 138: ...128 11 Product Full View...
Страница 139: ......
Страница 140: ......