92
Table 7.3 Setting Control Program Messages 11/13
Mnemonic Description
Format
RHS? Histogram
counter
readout
Returns count values of bar graphs divided into
ten and displayed on the screen as responses.
d1 to d2 are output in order of large threshold.
[Format]
RHS?
[Response]
d1 ~ d10: NR1 format
CHS Histogram
counter
clear
Corresponds to operation of [DATA] CLRH YES.
[Format]
CHS
BSZ?
Reads number of data groups in the measured
data buffer.
Format]
BSZ?
[Response]
d1: NR1 format
RBF?
Measured data buffer readout
d1 (Format designation: 0 ~ 1)
0: ASCII format
1: Binary format
All data in the data buffer is output continually
beginning oldest data
[Format]
RBF? d1
[Response]
If ASCII
If binary
Binary response data
with length
specification
CBF
Measured data buffer clear
Corresponds to operation of [DATA] CLRD ALL
[Format]
CBF
RDT?
Measured data query
d1 (Format designation: 0 ~ 2)
0: Basic format
1: Measured value only
2: Comparison result only
Most recent measured data is read out. This
command is used to read data when
measurement is made in the Internal Trigger
mode. Needs not be used in manual trigger,
external trigger and sequential measurement.
[Note]
See 7.14 “Using Program Message.”
[Format]
RDT? d1
d1: NR1 format
[Response]
(See 7.3)
ERR?
Error information clear
d1 (Error information: 0 ~ 255)
See Table 7.7 and Fig. 7.7
.
[Format]
ERR?
[Response]
d1: NR1 format
Содержание DSM-8542
Страница 1: ...DIGITAL SUPER Instruction Manual DSM 8542 July 2013 Revised edition 1 DS8542A981 01 13 07H MEGOHMMETER...
Страница 2: ......
Страница 21: ...11 3 2 Block Diagram of DSM 8542...
Страница 89: ...79 Syntax Chart Fig 7 1 Message Syntax Chart...
Страница 90: ...80 Command program header Fig 7 2 Program Header Syntax Chart...
Страница 91: ...81 Fig 7 3 Syntax Chart of Data Part...
Страница 138: ...128 11 Product Full View...
Страница 139: ......
Страница 140: ......