87
Table 7.3 Setting Control Program Messages 6/13
Mnemonic Description
Format
SEQ Sequential
mode
setting
d1 (Mode: 0 ~ 1)
0: OFF 1: ON
d2 (Program No.: 0 ~ 9)
No. of program executed when sequential mode
is turned ON
d3 (Discharging time before measurement)
(0.0 ~ 999.9s)
d4 (Charging time)
(0.0 ~ 999.9s)
d5 (Measuring time)
(0.0 ~ 999.9s)
d6 (Discharging time after measurement)
(0.0 ~ 999.9s)
[Note] Sequential mode: Subsequent parameters
are effective even on OFF.
(Retained as current set value)
[Format]
SEQ d1
d2, d3, d4, d5, d6
d1: NR1 format
d2: NR1 format
d3: NR2 format
d4: NR2 format
d5: NR2 format
d6: NR2 format
SEQ?
Same as set items of query SEQ command
message in sequential mode. Returns current
set data.
d1 (Mode: 0 ~ 1)
d2 (Program No.: 0 ~ 9)
d3 (Discharging time before measurement)
d4 (Charging time)
d5 (Measuring time)
d6 (Discharging time after measurement)
[Format]
SEQ?
[Response]
d1, d2, d3, d4, d5, d6
DEV
Deviation value display mode setting
d1 (Mode: 0 ~ 2)
0: OFF
1: DEV
2: PAR
d2 (Reference deviation value)
(-9.999E+30 ~ 9.999E+30)
[Note] Mode: d2 is enabled even when OFF.
(Retained as current set value)
[Format]
DEV d1, d2
d1: NR1 format
d2: NR3 format
DEV?
Query response in deviation value display mode
is same as setting.
Format]
DEV?
[Response]
d1, d2
Содержание DSM-8542
Страница 1: ...DIGITAL SUPER Instruction Manual DSM 8542 July 2013 Revised edition 1 DS8542A981 01 13 07H MEGOHMMETER...
Страница 2: ......
Страница 21: ...11 3 2 Block Diagram of DSM 8542...
Страница 89: ...79 Syntax Chart Fig 7 1 Message Syntax Chart...
Страница 90: ...80 Command program header Fig 7 2 Program Header Syntax Chart...
Страница 91: ...81 Fig 7 3 Syntax Chart of Data Part...
Страница 138: ...128 11 Product Full View...
Страница 139: ......
Страница 140: ......