Operating method
Using with the main 3194 (factory-fitted option)
However, it is necessary to change to the 3194 version.
Operating method
Applying as to the specifications of the main 3194 product
Temperature
coefficient
±
0.03%f.s./
℃
Analog output
±
5 Vf.s. Output accuracy = (Measurement accuracy)
±
0.2%f.s.
Response
FAST (0.1 s) /MID
(
0.8 s
)
/SLOW
(
5.0 s
)
(The analog output response time is the time such that for
an input change from 0% to 90% and 100% to 10% of the
nominal range value, the value is /- 1% of the final
stabilized value.)
Standards
Applying
Safety:
EN61010-1:1993+A2:1995 (using with the main 3194)
Pollution Degree 2, overvoltage category I
(anticipated transient overvoltage 330 V)
Chapter 5
Specifications
Configuration
There are two input connectors on chA. Each one corresponds to "strain
gauge input" and to "DC voltage input." On the 3194, it is necessary to set
which input connector to use. There is one input connector on chB. It
corresponds to "DC input" and "pulse input."
On the 3194, it is necessary to set the type of input.