8
_____________________________________________________________________________________________
3.4 Setting the Pulse
______________________________________________________________________________________________
F1
F2
F1
F2
F3
F4
Display value (rpm) =
Number of pulses for each
rotation of the motor
Measurement frequency
×
60
NOTE
Setting a pulse input signal to channel B unconditionally assigns the frequency
measurement function of the 3194 to fc. This ends any previous source set for fc.
The frequency range of fc is set to auto-ranging. To fix the frequency range, set
the range on "
FREQ/OUTPUT
" page on STATUS screen.
3.3 Setting the Products
3.4 Setting the Pulse
Input DC voltage values can be displayed together with a unit designation.
1. Display the "
EXT UNIT
" page on the STATUS
screen.
2. Using the
CURSOR
keys, move the cursor to"
UNIT
."
3. Press
F1
(
↑
) or
F2
(
↓
) to set the units.
Unit for channel A
(space)
⇔
V
⇔
N
・
m
⇔
mN
・
m
⇔
kN
・
m
⇔
kgf
・
m
⇔
kgf
・
cm
Unit for channel B
(space)
⇔
V
⇔
rpm
⇔
N
・
m
⇔
mN
・
m
⇔
kN
・
m
⇔
kgf
・
m
⇔
kgf
・
cm
When the voltage range for channel B is set to
PULSE
Hz
⇔
rpm
By selecting a pulse input to channel B, the 3194 frequency measurement
function can be used to measure a frequency. By setting the unit designation
to "rpm" this can be used to directly display the rotation speed of a motor or
other device.
1. Display the "
EXT UNIT
" page on the STATUS
screen.
2. Using the
CURSOR
keys, move the cursor
to"
VOLTAGE RANGE
" for channel B, and press
F4
(
PULSE
)
If the unit designation is set to "
rpm
", it is
necessary to set the number of pulses input per
rotation. In this case, set the Scaling item to the
number of pulses for each rotation of the motor.
Содержание 9603-01
Страница 1: ...INSTRUCTION MANUAL For 9603 01 EXTERNAL SIGNAL INPUT UNIT...
Страница 2: ......
Страница 4: ......