5
_____________________________________________________________________
Product Outline
______________________________________________________________
Maximum 400 points partitioned
at complete cycle
250
µ
s
Interval
Records instantaneous value
Interval
One second interval
Recording starts
Records average value
1.3 Measurement Value Recording Modes
Effective value calculation
To calculate effective value, measurement signal is sampled
every 250
µ
s and operation is carried out from data with
maximum 400 points to seek effective value.
Recording instantaneous value
To record instantaneous value, calculate effective value only
once per set interval and record in memory.
Recording average value
To record average value, calculate effective value only once
per second and record average value of all data within
interval in memory.
After recording starts, average value within interval is
recorded, so the initial data is not recorded immediately
after recording starts but is recorded from the following
interval.