11
_____________________________________________________________________
Set Up
______________________________________________________________
WARNING
Maximum input current for 9650 is 100 Arms, 9651 is
500 Arms. Do not exceed the continuous maximum
input current.
To avoid short circuit and the risk of accidental injury or
death, working voltage must be under AC300 Vrms (CAT
III) for 9650 and under AC600 Vrms (CAT III) for 9651.
When the clamp sensor is opened, do not allow the
metal part of the clamp to touch any exposed metal, or
to short between two lines, and do not use over bare
conductors.
When connecting with 9650 CLAMP ON SENSOR and
3636-20 measurement range is set at 50 A, do not
exceed 50 Arms.
To avoid damaging the unit, do not use any other
sensors except 9650 or 9651 CLAMP ON SENSOR as
sensor connector.
9650 or 9651
CLAMP ON SENSOR
9650
9651
Carefully read 9650 and 9651 CLAMP ON
SENSOR instruction manuals before use.
Improper measurement method may result in
accidental injury, death or damage in the unit.
2.4 Connecting Sensor/Cable
Connecting clamp on sensor
3636-20 CLAMP LOGGER enables 2 channels of data
measurement when connected with 9650 or 9651 CLAMP
ON SENSOR.
When measuring, affix clamp to one conductor only.