16
_____________________________________________________________________
Settings
______________________________________________________________
Average value recording mode records average
value within interval, so the initial data is not
recorded immediately after recording starts but
is recorded from the following interval.
For example, when measurement started at
12:00 at 1 minute intervals, the first data is data
recorded when 1 minute has passed at 12:01.
Alarm signal output terminal is an open drain
output. During signal output, signal side and
GND side are internally grounded and any other
time, it remains open. When external power
supply is supplied, alarm signal can control
relay or sequencer. Maximum rate: 30 V, 200
mA (allowable loss 200 mW)
9. Recording mode setting
With 3636-20 stand-alone, or connected together with
COMMUNICATION BASE and PC, setting is available.
Two recording mode options are instantaneous value
recording and average value recording (
AV
displayed).
At factory shipment setting, average value recording is
selected.
10. Alarm setting
Alarm setting is available when 3636-20,
COMMUNICATION BASE and personal computer are
connected.
Designate upper limit and lower limit value of measurement
value with personal computer to output alarm signal when
measurement value goes outside of set range. Two channels
pass individual judgment.
During signal output,
AL
appears on the display.