22
―――――――――――――――――――――――――――
Chapter 2
Measurement Procedure
――――――――――――――――――――――――
U
I
φ
t
2.4.1 1
φ
P Meter, 1
φ
PF Meter and 3
φ
PF Meter
[1
φ
P Meter]
Displays active power P once about every second
(once about every three seconds in SLOW mode).
The meter calculates apparent power S, reactive
power Q, and power factor COS
φ
from active
power P, voltage U, and current I. (See 3.3,
"Operation Expressions.")
[1
φ
PF Meter and 3
φ
PF Meter]
The phase angle is measured at the zero-cross point
of voltage U and current I as shown below. The
meter calculates three-phase active power P, three-
phase apparent power S, three-phase reactive power
Q, and reactive factor SIN
φ
, and power factor COS
φ
from the phase angle
φ
, voltage U, and current I.
(See 3.3, "Operation Expressions.")
For an inverter or thyristor with distorted input
waveforms, or waveforms with noise superimposed,
the meters may not display accurate values or even
be able to measure at all.
Three-phase active power P is calculated on the 3
φ
PF meter under balanced load conditions. Accurate
measurements cannot be conducted under an
unbalanced load.
Содержание 3286-20
Страница 1: ...EN 3286 20 CLAMP ON POWER HiTESTER Instruction Manual Apr 2015 Revised edition 13 3286B981 13 15 04H...
Страница 17: ...xiv Organization of This Manual...
Страница 20: ...3 Chapter 1 Product Outline Top and Side View 1 2 3 4 5 6 9 8 13 14 10 11 15 16 12 7 17 1 3 Parts and Functions...
Страница 29: ...12 Chapter 1 Product Outline...
Страница 55: ...38 Chapter 2 Measurement Procedure...
Страница 69: ...52 Chapter 3 Specifications...
Страница 73: ...56 Chapter 5 Attaching The Hand Strap...
Страница 75: ...58 Chapter 6 Storage in Carrying Case...